English Operating Manual Refrigerated/Heating Circulators air-cooled F25-HL F32-HL F33-HL FP35-HL FP40-HL FP45-HL FP50-HL water-cooled FPW50-HL Translation of the Original Operating Manual 1.951.2809-V2 19512809-V2.doc 06/13 JULABO GmbH 77960 Seelbach / Germany Tel. +49 (0) 7823 / 51-0 Fax +49 (0) 7823 / 24 91 info@julabo.de www.julabo.de 22.07.
Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been designed to help you gain an understanding of the operation and possible applications of our circulators. For optimal utilization of all functions, we recommend that you thoroughly study this manual prior to beginning operation.
HL TABLE OF CONTENTS Operating manual .................................................................................................................. 5 1. Intended use ................................................................................................................... 5 1.1. Description ................................................................................................................. 5 2. Operator responsibility – Safety recommendations ......................................
9.4.2. Keypad control or setpoint setting via the analog input ........................................ 50 9.4.3. Autostart ............................................................................................................... 51 9.4.4. Off-Mode .............................................................................................................. 51 9.4.5. ActVar - actuating variable ................................................................................... 52 9.4.6.
HL Operating manual 1. Intended use JULABO circulators have been designed to control the temperature of specific fluids in a bath tank. The units feature pump connections for temperature control of external systems (loop circuit). JULABO circulators are not suitable for direct temperature control of foods, semiluxury foods and tobacco, or pharmaceutical and medical products. Direct temperature control means unprotected contact of the object with the bath medium (bath fluid). 1.1.
Operator responsibility – Safety recommendations 2. Operator responsibility – Safety recommendations The products of JULABO ensure safe operation when installed, operated, and maintained according to common safety regulations. This section explains the potential dangers that may arise when operating the circulator and also specifies the most important safety precautions to preclude these dangers as far as possible. The operator is responsible for the qualification of the personnel operating the units.
HL Use: The bath can be filled with flammable materials. Fire hazard! There might be chemical dangers depending on the bath medium used. Observe all warnings for the used materials (bath fluids) and the respective instructions (safety data sheets). Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in well ventilated areas. Only use recommended materials (bath fluids). Only use non-acid and non corroding materials.
Operator responsibility – Safety recommendations 2.1. Disposal The circulator contains a back-up battery that supplies voltage to memory chips when the unit is switched off. Do not dispose of the battery with household waste! Depending on battery regulations in your country, you might be obliged to give back used or defect batteries to gathering places. The product may be used with oil as bath fluid. These oils fully or partially consist of mineral oil or synthetic oil.
HL 2.2. EC Conformity The products described in the operating instructions conform to the requirements of the following European guidelines: Directive of the European Parliament and of the Council on the approximation of the laws of the Member States relating to machinery. EMC guideline with respect to legal harmonization of the member countries concerning electromagnetic compatibility. JULABO GmbH Eisenbahnstr. 45 77960 Seelbach / Germany 2.3.
Operator responsibility – Safety recommendations 2.4. Technical specifications F25-HL F32-HL Working temperature range °C -28 ... 200 -35 ... 200 Temperature stability °C ±0,01 ±0,01 Cooling capacity Medium: ethanol Refrigerant °C kW +20 0 -20 0.26 0.2 0.06 R134a +20 0 -20 -30 0.45 0.39 0.15 0.06 R134a (WxDxH) cm 23x42x64 31x42x64 (WxL) cm 12x14 18x12 Bath depth cm 14 15 Filling volume liters 3 ... 4.5 5.5 ...
HL FP35-HL FP40-HL Working temperature range °C -35 ... 200 -40 ... 200 Temperature stability °C ±0.01 ±0.01 Cooling capacity Medium: ethanol Refrigerant °C kW +20 0 -20 -30 0.45 0.39 0.15 0.05 +20 0 -20 -40 0.68 0.5 0.32 0.04 R134a R134a (WxDxH) cm 31x42x66 36x46x71 (WxL) cm 18x12 23x14 Bath depth cm ----- 20 Filling volume liters 2.5 9 ...
Operator responsibility – Safety recommendations Working temperature range Temperature stability Cooling capacity Medium: ethanol Refrigerant Overall dimensions FPW50-HL -50 ... 200 ±0.01 +20 0 -20 -40 0.9 0.8 0.5 0.16 °C °C °C kW R404A / R507 (WxDxH) cm 42x49x72 (WxL) cm 18x12 Bath depth cm 20 Filling volume liters 5.5 ...
HL Electrical connections: External alarm device 24-0 V DC / max. 25 mA Computer interface RS232 or RS485 External Pt100 sensor Optional for HL, SL (Order No.
Operator responsibility – Safety recommendations 2.5. Cooling water connection Cooling water pressure (IN / OUT ) Difference pressure (IN - OUT ) Cooling water temperature max. 6 bar 3.
HL Operating instructions 3. 3.1. Safety notes for the user Explanation of safety notes In addition to the safety warnings listed, warnings are posted throughout the operating manual. These warnings are designated by an exclamation mark inside an equilateral triangle. “Warning of a dangerous situation (Attention! Please follow the documentation).” The danger is classified using a signal word. Read and follow these important instructions for averting dangers.
Safety notes for the user 3.3. Safety recommendations Follow the safety instructions to avoid personal injury and property damage. Also, the valid safety instructions for workplaces must be followed. Only connect the unit to a power socket with an earthing contact (PE – protective earth)! The power supply plug serves as a safe disconnecting device from the line and must always be easily accessible.
HL Caution: The temperature controlling i.e. of fluids in a reactor constitutes normal circulator practice. We do not know which substances are contained within these vessels. Many substances are: inflammable, easily ignited or explosive hazardous to health environmentally unsafe i.e.: dangerous The user alone is responsible for the handling of these substances! The following questions shall help to recognize possible dangers and to reduce the risks to a minimum.
Operating controls and functional elements 4.
HL 1a 1b 2 Mains power switch, illuminated for circulator Mains power switch, illuminated for cooling machine VFD COMFORT-DISPLAY Header: Control indicators Line 1: Actual value internal or external The display is depending on the selected control mode in the menu > Control < (internal or external). Line 2: Working temp. setpoint, constantly S xxx.
Operating controls and functional elements Navigation aid in MENU - Window / : Keys Selection of menu items / parameters - Setting in line 3: Actual value / parameter Orientation aid in MENU - Window Line 1 - name with allocation to key Safety values T – Setpoint 1 Main menu Level 1 2 Submenu Level 2 Example: CONFIGURATION 3 Submenu Level 3 4 4.1 Keypad 1 Navigation keys 1. Key: >OK< Start / Stop (pump / heater ) 2.
HL 5 Keypad 2 Numeric keypad: numerals 0 to 9 .. minus / decimal point Backspace key on LCD DIALOG-DISPLAY. 6 Adjustable excess temperature protection according to IEC 61010-2-010 Rear view 7 Socket for external measurement and control sensor or external setpoint programming ext Pt100 8 Interface RS232 / RS485: remote control via personal computer SERIAL 9 9a Control cable of JULABO refrigerated circulator 9b Socket: Socket: output for alarm messages Option: Electronic module Order No.
Operating controls and functional elements 17a 17b Mains power cable with plug for circulator 18 Built-in mains outlet for connection of circulator (230 V / 16 A) 19 Drain screw with drain connection 20 Venting grid, removable 21 22 Pump connections 23 Mains power cable with plug cooling machine suction pump Selector dial for cooling machine (only F25) Position "1" for operation with HL circulator. 24 Cooling water OUTLET and INLET.
HL 5. 5.1. Preparations Installation Place the unit on an even surface on a pad made of non-flammable material. Cooling machine, pump motor and electronics produce intrinsic heat that is dissipated via the venting openings! Never cover these openings! Keep at least 20 cm of open space on the front and rear venting grids.
Preparations Ferrous water can cause corrosion - even on stainless steel. Chloric water can cause pitting corrosion. Distilled and deionized water is unsuitable. Their special properties cause corrosion in the bath, even in stainless steel.
HL 5.3. Temperature application to external systems Caution: Securely attach all tubing to prevent slipping. If the circulator is operated without external system, close the pump connector (22) with the cap nut. Temperature application to external, closed systems The circulator is used for temperature application to external, closed systems (loop circuit) with simultaneous temperature application in the circulator bath.
Preparations Backflow safety device ext Pt100 ALARM STAND-BY SERIAL If the liquid levels in the circulator bath and the external system are at different heights, overflowing must be prevented after the power has been turned off. REG+ E-PROG 13 T16A 14 T16A Flood hazard! T1.25A For this reason, solenoid valves for loop circuit or shut-off valves can be integrated in the loop circuit. Connect the valve to the connector (13).
HL Safety recommendations Employ suitable connecting tubing. Make sure that the tubing is securely attached. Avoid sharp bends in the tubing, and maintain a sufficient distance from surrounding walls. Regularly check the tubing for material defects (e.g. for cracks). Preventive maintenance: Replace the tubing from time to time. 5.4. Filling / draining Notice: Pay attention to the thermal expansion of bath oil during heating to avoid overflowing of the liquid.
Operating procedures 6. 6.1. Operating procedures Power connection Caution: Only connect the unit to a power socket with earthing contact (PE – protective earth)! The power supply plug serves as safe disconnecting device from the line and must be always easily accessible. Never operate equipment with damaged mains power cables. Regularly check the mains power cables for material defects (e.g. for cracks).
HL 6.2.2. Switching on the cooling machine Switching on: Switch on the cooling machine using the switch (1b) . Control of the cooling machine: With the mains switch (1b) turned on, the circulator automatically switches the cooling machine off and on. It is switched off if: - the current working temperature is increased by > 30 °C (cooling is not required). - the heater operates at full power (> 800 W) for longer than 5 minutes.
Setting of temperatures 7. Setting of temperatures Press the key to call up the menu for temperature selection. 3 different working temperatures can be adjusted. Their values are freely selectable within the operating temperature range. The temperatures can be set in start or stop mode. menu Setting of working temperature in the Factory settings: 1. Press the key . The value flashes 2. Select SETPOINT 1 or 2 or 3 using the key 3. Confirm by pressing the or . key.
HL Warning: The excess temperature protection must be set at least 25 °C below the fire point of the bath fluid used! In case of wrong setting there is a fire hazard! We disclaim all liability for damage caused by wrong settings! SafeTemp: Setting range: 20 °C ... 320 °C Rough setting can be effected by using the temperature scale. Settings: SafeTemp: Type: Press the key the menu item and by pressing the key or and confirm the by pressing the key select . For setting proceed as follows.
Safety installations, warning functions The following error message appears on the VFD COMFORT-DISPLAY: for help on the LCD Press Depending on the setting of >ControlType - internal or external< the actual temperature values for both sensors are shown/indicated on both displays. Examples: I 94.06 and Ext. : xx.
HL Setting >WARNING< A mere warning function with optical and audible warning signal (interval tone) A message appears on the VFD COMFORT-DISPLAY: or OverTemp SubTemp The LCD –DISPLAY shows the message: Setting >ALARM< Temperature limit with shutdown of heater and circulating pump.
Safety installations, warning functions 8.0.1. Early warning system, low level protection This low level protection is independent of the control circuit and is divided into two sections: DBGM: 20306059.8 1. Switch in stage 1 recognizes a critical fluid level . An audible warning (interval tone) sounds and a message appears on the VFD COMFORT-DISPLAY. Refill bath fluid! 2. Switch in stage 2 recognizes a low fluid level .
HL 9. Menu functions The term „menu functions“ refers to adjustments such as Menu level 1 Start program Page 37 This menu will start a previously set program. Administration and creation of programs Page 42 Electronically adjustable pump capacity Page 46 The capacity of the circulating pump is set by adjusting the motor speed.
Menu functions Menu level 1 Adjustable interface parameters BAUDRATE, PARITY, HANDSHAKE Page 60 ATC - Absolute Temperature Calibration, Sensor calibration internal sensor, Sensor calibration external sensor 3-point calibration Page 61 Limitations of temperature and capacity Page 68 Setpoint Max / Min - Maximum and minimum setpoint Heating Max – Set maximum heating Cooling Max – Set maximum cooling Internal Max / Min – Limitation of the temperature range Upper band / Lower band – Band limit Analog inputs
HL 9.1. Start of a program The start menu of the integrated programmer allows one of six previously stored temperature programs to be called up and started. The profiles are started manually or via the integrated timer. Requirements: 1. Create a program. (refer to next chapter) 2. Return to the Start-MENU and confirm the desired setting of each MENU item with the key 3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be started by the internal timer.
Menu functions Level 2 Parameter /Value or Set desired parameters with and . Confirm >now< with the key and the program will start immediately or start at the set time under parameter (time ). Set time in the example below: 09. August 2009, 11:15 hrs Submenu TIMER If the parameter time is selected, a further submenu opens for setting the start time. Set the time for the start of the program in the submenu >3 TIMER<.
HL 9.1.1. Status at the end of a program Program end StdBy , PSetp, Setp 1, Setp 2 , Setp 3 This parameter is used to determine if the unit switches to the -OFFcondition at the end of a profile or if temperature control is continued, and the working temperature setpoint that is to be used if temperature control is continued. StdBy – If the end of the profile is reached, the circulator switches to the -OFF- condition.
Menu functions 9.1.3. Editing after Start The programmer allows a running program to be edited. Initiation and exit of the editing function: After the start enter the submenu „3 EDITOR“.by pressing the key See chapter „program administration, creation“ page 42 Exit the editing modus anytime by pressing the key . Particularity for changes in a section that is currently in process.
HL 9.1.5. Interruption after a power failure It is possible to operate the programmer safely in case of a power failure. Important: For this, >Autostart< needs to be set in the >Configuration< menu. If the AUTOSTART function is activated, the programmer starts again at a point approx. 20 seconds before the interruption took place. However, an uncontrolled change of the bath temperature has occurred.
Menu functions 9.2. Program administration, creation The integrated programmer permits fast and easy programming of setpoint temperature sequences. This temperature sequence is called program. A program is composed of individual sections (Step). The sections are defined by duration (t) or Gradient (°/t) and target temperature. The target temperature is the setpoint (SP), which is achieved at the end of a section. The programmer calculates the temperature ramp from the difference in time and temperature.
HL Editing PrgNr: Set the program you wish to create. 0 ... 5 SP : Setpoint TI : GRD: Time [hh:mm] or Gradient [°/t] = [°/min] Change type of program in MENU >configuration<, menu item >Programmtyp/program type< . ST: Step Press the 0 … 59 key if a parameter is to be retained. (prior to OK) Level 2 Example: Program (PrgNr.) 2 is to have a target temperature with a set point (SW) of 100.00 °C in section t (ABS) 3. Duration (ZI) 1 hour. Select the program number. Set the desired >PrgNr.
Menu functions Delete It is possible to delete an entire program or consecutive sections Delete PrgNr: Set the program you wish to delete 0 ... 5 Delete from Step X to Step X - (0 … 59) Press the key if a parameter is to be retained. (prior to OK) Level 2 Submenu >Delete< Set menu item >Delete PrgNr.< by pressing and . Set desired PrgNr with and . Example: PrgNr. 0 Level 3 44 Input window: delete sections. Select menu items > from Step< and > to Step< one and .
HL Print Every program can be printed via the serial interface for control and documentation purposes. Press the key if a parameter is to be retained. (prior to OK) Level 2 Submenu >Print< Set menu item >Print PrgNr.< by pressing and . Set desired >PrgNr.< by pressing and . Example: 4 Level 3 Confirm Confirm >Print?< by pressing Printing and end of printing are confirmed by successive messages on the LCD display as shown on the left. > Print< by pressing . .
Menu functions 9.3. Setting the pump pressure The capacity of the circulating pump is set by adjusting the motor speed Press the key if a parameter is to be retained. (prior to OK) Submenu >Pump< Level 2 The set pump stage is displayed. Example: 1 Confirm by pressing Set required pump stage by pressing and . . Example: 3 Example: Adjusted Adjustable pump capacity Effective Illuminated display: stage 1 ...
HL 9.4. Configuration Level 2 Press the key if a parameter is to be retained. (prior to OK) Parameter level Remote control via serial interface. Possible parameters: off / on or Switch on and off remote control by and pressing First set the >type< of interface (RS232 or RS485) in the MENU >INTERFACE SERIAL< and then switch >Remote< to >on<. Control display in the top line for Remote Connect SERIAL with PC.
Menu functions Level 2 Parameter level Pump-Mode Possible parameters: P. on / P. off Switch on and off >Off-Mode by pressing and P.on continuous operation of circulating pump P.off circulating pump is linked to Start/Stop ActVar - actuating variable Possible parameters: control, Eprog, serial Set desired parameters with and .
HL Level 2 Parameter level Reset Level 2 Level 3 Return to factory settings by pressing >Reset?< returns all set values to the factory setting except for date and time. A RESET can be effected only in the – OFF- mode. During the message >Running < all parameters are reset to factory settings. Parameter / Value Possible parameters: Time >hour:min Date >Day:Mon Year Succesively set time, date and year using the numeric keypad and confirm by pressing .
Menu functions 9.4.1. Remote control via the serial interface Factory setting: off Setpoint is set via the serial RS232/RS485 interface through a PC or superordinated data system. In the header of the VFD COMFORT-DISPLAY, an “R” illuminates. It indicates that remote control mode is active. The selected setting is shown on the LCD DIALOG-DISPLAY.
HL 9.4.3. Autostart Possible parameters: on - Autostart on off - Autostart off (Factory setting) Note: The circulator has been configured and delivered by JULABO according to N.A.M.U.R. recommendations. This means for the start mode that the unit must enter a safe operating state after a power failure (nonautomatic start mode). This safe operating state is indicated by “OFF” or “R OFF” on the VFD-COMFORT-DISPLAY.
Menu functions 9.4.5. ActVar - actuating variable The variable (ACTuating VARiable) corresponds to the extent to which the heater or cooling unit of the circulator is controlled. Heat or cold is applied to the bath according to this variable. If this happens with the control electronics of the circulator, called > control < in this particular case, the bath temperature is exactly heated and maintained constant at the adjusted setpoint.
HL 9.4.6. Setting of clock and date The internal real time clock allows starting a program any time. The clock is set to the local mean time (MEZ) at the factory. 9.4.7. If the unit is operated in a different time zone, the clock can be adjusted in this menu. Change summer/winter time in this menu Language There are two options for the language of the LCD DIALOG-DISPLAY: German and English. Possible parameters: deutsch / english 9.4.8. Unit Temperature values can be displayed in °C or °F.
Menu functions 9.5. Control The circulator is designed for internal and external temperature control. This can be switched in this submenu. Only the set of parameters corresponding to the particular setting is displayed. Control internal Control external For external temperature control and measurement connect a Pt100 external sensor to the socket at the rear of the circulator. Press the key if a parameter is to be retained.
HL Control Type internal Parameter-Level The value flashes. Set by using the numeric keypad and confirm by pressing . Example: Setting range: 0.1 … 99.9 K The value flashes. Set by using the numeric keypad and confirm by pressing . Setting range: 3 … 9999 s The value flashes. Set by using the numeric . keypad and confirm by pressing Setting range: 0 … 999 s Control Type external Parameter-Level The value flashes. Set by using the numeric keypad and confirm by pressing .
Menu functions 9.5.1. Control internal / external The control type can be adjusted in the -OFF- mode only. Possible parameters: intern internal temperature control extern external temperature control with external Pt100 sensor IMPORTANT: Additional measures for external temperature control. ext. Pt100 Connect a Pt100 sensor to the socket on the rear of the circulator. Sensor calibration of the external Pt100 sensor is performed in the >ATCalibration< menu.
HL 9.5.2. Selftuning When performing a selftuning for the controlled system (temperature application system), the control parameters Xp, Tn and Tv are automatically determined and stored. Possible parameters: off - no selftuning The control parameters ascertained during the last identification are used for control purposes.
Menu functions 9.5.4. CoSpeed - extern CoSpeed: This parameter affects the temperature pattern only in case of external control. Possible parameters: 0 ... 5 °C Int. temperature Ext. temperature 5 3 0 Setpoint t 9.5.5. During selftuning, the control parameters Xp, Tn, and Tv of a controlled system are automatically determined and stored. The time required for tuning may vary depending on the controlled system. This controller design allows protection of sensitive objects requiring temperature control.
HL Optimization instructions for the PID control parameters: optimum setting The heat-up curve reveals inappropriate control settings.
Menu functions 9.6. Serial Interface Factory settings: 4800 Baud even hardware handshake For communication between the circulator and a PC or a superordinated process system, the interface parameters of both units must be identical. The adjustment is usually carried out one time only in the >Interface serial< menu. RS232 RS485 Level 2 Parameter-level Press the key if a parameter is to be retained.
HL 9.7. ATC Absolute Temperature Calibration, 3-point calibration ATC is used to compensate for a temperature difference between the circulator and a defined measuring point in the bath tank that may develop for physical reasons. ATC internal external Principle: For ATC calibration, in steady state the bath temperature at the location of the temperature sensor (CT) is determined at the respective adjusted working temperature.
Menu functions Press the key if a parameter is to be retained. (prior to OK) Level 2 Parameter-Level The parameter flashes, switch by pressing and or The parameter flashes, switch by pressing and or >no< Carry out an ATC calibration >yes< return to standard operation after calibration. The parameter flashes, switch by pressing and A >1-point<, >2-point< or >3-point< calibration can be carried out.
HL The value >TmpValue< is only indicated In addition the measured temperature value >CalValue X< is saved during the next step. The value flashes. Set by using the numeric keypad and confirm by pressing . If only a 2-point calibration is carried out, the following menu items are not indicated anymore The value >TmpValue< is only indicated. In addition the measured temperature value >CalValue X< is saved during the next step.. The value flashes.
Menu functions 9.7.1. ATC Fühler - intern / extern In the first submenu the ATC function is set for the >intern< internal or the >extern< external temperature sensor. Calibration can be carried out for the internal temperature sensor and for the external temperature sensor connected to the socket „ext. Pt100“. The circulator is able to save both parameter sets. However only the one which has been set under menu item > sensor < is displayed. 9.7.2.
HL 9.7.4. Example: 3-point calibration for internal control In the temperature range from 80 °C to 160 °C the calibration curve of the temperature sensor (TT) is to be adjusted to the actual temperatures at measuring point (CT). 1. Set circulator to internal control: The type of control can be set only in the –OFF- mode. 2. Set working temperature setpoint See standard display line 1 80.00 °C 120.00 °C 160.00 °C Example 80.00 °C Press the keys and and confirm by pressing .
Menu functions The value >TmpValue< is only indicated. In addition the measured temperature value >CalValue X< is saved during the next step. The value flashes. Set by using the numeric keypad and confirm by pressing . Return to 2. Set working temperature value: 120.00 °C The value >TmpValue< is only indicated. In addition the measured temperature value >CalValue X< is saved during the next step. The value flashes. Set by using the numeric keypad and confirm by pressing .
HL 9.8. Limits The >Limits< menu allows the minimum and maximum values to be set for all important setting ranges and capacity variables. Control external Control internal Press the key if a parameter is to be retained. (prior to OK). Level 2 Parameter-level The parameter flashes. Set by using the numeric keypad and confirm by pressing .. The parameter flashes. Set by using the numeric keypad and confirm by pressing .. The parameter flashes.
Menu functions The parameter flashes. Set by using the numeric keypad and confirm by pressing . The parameter flashes. Set by using the numeric keypad and confirm by pressing . The parameter flashes. Set by using the numeric keypad and confirm by pressing . The parameter flashes. Set by using the numeric keypad and confirm by pressing . 9.8.1.
HL 9.8.2. Limits for external control INTERN MAX / MIN Restriction for the temperature range of the internal bath. Setting range: -94,9 °C ... +300,0 °C The limits IntMax and IntMin are only active in external control. IntMax and IntMin determine fixed limits for the temperature within the internal bath. The temperature controller cannot exceed these limits even if it would be necessary for achieving the temperature in an external system.
Menu functions 9.9. Analog inputs/outputs In order to use the analog inputs and outputs, the circulator must be ALARM STAND-BY REG+ E-PROG equipped with the electronic module, which is available as option. Order No. 8 900 100 Electronic module This submenu enables the input and output values to be set for the programmer input and the temperature recorder outputs of the REG+EPROG socket. The >STAND-BY< input and the >ALARM< output are configurable.
HL Level 2 Level 3 Parameter-level Define the output value for channel 1 Possible parameters: ActInt, ActExt, Power, Setpoint The parameter flashes, switch by pressing and Select the scale for channel 1. The value flashes. Set by using the numeric keypad and confirm by pressing . The value flashes. Set by using the numeric keypad and confirm by pressing . Define the output value for channel 2.
Menu functions Level 2 Level 3 Parameter-level Define the output value for channel 3 Possible parameters: ActInt, ActExt, Power, Setpoint The parameter flashes, switch by pressing and Select the scale for channel 3. The value flashes. Set by using the numeric keypad and confirm by pressing . The value flashes. Set by using the numeric keypad and confirm by pressing .
HL Level 2 Level 3 Parameter-level Possible parameters: voltage, current The parameter flashes, switch by pressing and (Parameter Level) The value flashes. Set by using the numeric keypad and confirm by pressing . The value flashes Set by using the numeric . keypad and confirm by pressing The value >setpoint< is only indicated.
Menu functions Level 2 Level 3 Parameter-level Possible parameters: normal, inverse The parameter flashes, switch by pressing and 9.9.1. Control connector Control connector (14): SV-Out >Backflow< or >C-Pulse< or >Refill< The control output (14) can be adapted to solenoid valves for different tasks. >Backflow< In the OFF condition, the control output (14) is not powered. See – Backflow safety device- on page 26 >C-Pulse< The control output (14) can be used to emit a cooling pulse.
HL 9.9.2. Outputs of the connector - REG+E-PROG Outputs of the connector 1. First define the desired output value for channels 1 to 3: ActInt internal actual temperature value (bath temperature) ActExt external actual temperature value (external sensor) Power periodic or intermittent heating or cooling Setpoint active setpoint temperature (setpoint 1, 2, 3/integr. programmer/external programmer) 2.
Menu functions 9.9.3. Input of the connector - REG+E-PROG E-PROG - Input Menu >Configuration< 1. 2. Menu >Inputs/Outputs< Setting is necessary if 1. the Setpoint is to be set via an external voltage or current source or programmer For this, in the menu >Configuration<, first set the menu item >Setpoint< to >Eprog<. 2. the heater variable should be controlled via an external control pulse. For this, in the menu >Configuration<, set the menu item > ActVar.< to >Eprog<. 3.
HL After returning the LCD display to standard display by pressing the , the temperature value adjusted and set on the external key voltage or current source is displayed in line 1 (Example: Eprog 50.00 C). This EPROG input enables the use of different voltage and current values as program parameters. °C “L Value” - Setting the Low value: (See below ) 1) Adjust and set the lowest desired value on the voltage or current source (Example A: 1 V). Wait approximately 30 seconds.
Menu functions 9.9.4. External Stand-by input Ext. StBy (External STAND-BY input) Stand-By input for external switch-off. Possible parameters: inactive - Stand-By input is ignored active - Stand-By input is active Activate the standby input: 1. Under menu item >Ext. StBy<, set the parameter to >active<. 2. Connect an external contact “AK” (e.g., for external switch-off) or an alarm contact of the superordinated system.
HL 9.9.5. Alarm-output Alarm output (10) (for external alarm signal) This socket is a potential-free change-over contact. With the adjustments in the menu >Inputs/Outputs<, all operating conditions can be signaled without having to change the pin assignments. Meaning of the terms under menu item >Function< : ALARM The circulator is in condition >StandBy< or >Alarm< For >Type: normal<, pins 2 and 3 are connected in any case according to the selected >Function<.
Troubleshooting guide / Error messages 10. Troubleshooting guide / Error messages Alarm with complete shutdown: If one of the following failures occur a complete, all-pole shutdown of the heater and circulating pump is effected. + “ lights up and a continuous signal sounds. „ The code for the cause of alarm is indicated on the VFD COMFORTDISPLAY. Alarm without shutdown: The code for the cause of alarm is indicated on the VFD COMFORTDISPLAY. The warning signal sounds in regular intervals.
HL Defect of the working or excess temperature sensor. The working temperature and excess temperature sensors report a temperature difference of more than 35 K. Other errors (I2C-BUS errors) Error in A/D converter Excess temperature sensor is defective. The safety temperature is below the working temperature setpoint. Set the safety temperature to a higher value. External control selected, but external Pt100 sensor is not connected or is defective. Cooling of the condenser is impaired.
Troubleshooting guide / Error messages Short-circuit in the control cable for the cooling machine during self-test. Cable of the excess temperature sensor disconnected or short-circuited. Ext. Pt100 sensor input without signal, but setpoint programming set to external Pt100. The early warning system for low level signals a critical fluid level. Replenish the bath tank with bath fluid.
HL 11. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected to the plug housing. The unit ensures safe operation if connecting cables with a maximum length of 3 m are used. The use of longer cables does not affect proper performance of the unit, however external interferences may have a negative impact on safe operation (e.g. cellular phones). 2 1 3 4 ext.
Electrical connections RS232 interface cable Circulator (9-pol) Pin 2 RxD Pin 3 TxD Pin 5 GND Pin 7 RTS Pin 8 CTS Accessories: PC (9-pol) Pin 3 TxD Pin 2 RxD Pin 5 GND Pin 8 CTS Pin 7 RTS Order No. Description 8 980 073 8 900 110 RS232 interface cable 9-pol./9-pol. , 2,5 m USB interface adapter cable / Control output The connector may only be used for control of a JULABO refrigerated circulator or JULABO MVS Solenoid valve controller for cooling water.
HL Stand-by input (for external switch-off) Pin assignment: AK 2 Pin 1 2 3 Signal not connected 5 V / DC 0V 3 1 STAND-BY Programmer input / temperature recorder output 3 2 4 1 5 6 REG+E-PROG Pin 1 Voltage output Channel 1 2 Voltage output Channel 2 3 GND for outputs 0V 4 Programmer input EPROG 5 Current output Channel 3 6 GND for Progammer 0V Signal 0 ... 10 V 0 ...
Remote control 12. Remote control 12.1. Setup for remote control 5 9 1 6 SERIAL 1. Check the interface parameters for both interfaces (on the circulator and PC) and make sure they match. In the >Interface< menu, set the menu item >Type< to >RS232< or >RS485<. 2. In the >Configuration< menu, set the menu item >Setpoint< to >RS232< or >RS485<. 3. Connect both units with an interface cable.
HL Important times for a command transmission: To ensure secure data transfer, the time gap between two commands should be at least 250 ms. The circulator automatically responds to an in command with a data string followed by an LF (Line Feed). The next command should be sent only after 10 ms. The out commands are valid only in remote control mode. Examples: When the RS485 interface is used, the three-digit instrument address precedes each command.
Remote control Command Parameter Response of circulator out_mode_03 0 Set external programmer input to voltage. Voltage 0 V ... 10 V out_mode_03 1 Set external programmer input to current. Current 0 mA ... 20 mA out_mode_04 0 Temperature control of internal bath. out_mode_04 1 External control with Pt100 sensor. out_mode_05 0 Stop the unit = R –OFF-. out_mode_05 1 Start the unit.
HL in commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of circulator version None Version number of the software (V X.xx) status none Status message, error message (see page 91) in_pv_00 none Current bath temperature. in_pv_01 none Heating power being used (%). in_pv_02 none Temperature value registered by the external Pt100 sensor. in_pv_03 none Temperature value registered by the safety sensor.
Remote control Command Parameter Response of circulator in_par_12 none Tv control parameter of the cascade controller. in_par_13 none Adjusted maximum internal temperature of the cascade controller. in_par_14 none Adjusted minimum internal temperature of the cascade controller.
HL 12.4. Status messages Status messages Description 00 MANUAL STOP Circulator in „OFF“ state. 01 MANUAL START Circulator in keypad control mode. 02 REMOTE STOP Circulator in „r OFF“ state. 03 REMOTE START Circulator in remote control mode. 12.5. Error messages Error messages Description -01 LOW LEVEL ALARM Low liquid level alarm -02 REFRIGERATOR ALARM Control cable of the refrigerated circulator or MVS solenoid valve controller short-circuited or disconnected.
Installation of electronic module with analog connectors Error messages Description -20 WARNING: CLEAN CONDENSOR OR CHECK COOLING WATER CIRCUIT OF REFRIGERATOR Cooling of the condenser is impaired. Clean air-cooled condenser. Check the flow rate and cooling water temperature for a water-cooled condenser. -21 WARNING: COMPRESSOR STAGE 1 DOES NOT WORK First-stage compressor does not work. -22 WARNING: COMPRESSOR STAGE 2 DOES NOT WORK Second-stage compressor does not work.
HL 14. JULABO Service – Online remote diagnosis JULABO circulators of the HighTech series are equipped with a black box. This box is implemented in the controller and records all significant data for the last 30 minutes. In case of a failure, this data can be read out from the unit by using special software. This software is available as a free download from www.julabo.de \ EasyBlackBox. Installation is easy and is performed step by step. Please observe the instructions.
Cleaning / repairing the unit 15. Cleaning / repairing the unit Caution: Always turn off the unit and disconnect the mains cable from the power source before cleaning the unit. Prevent humidity from entering into the circulator. Electrical connections and any other work must be performed by qualified personnel only. To maintain the full cooling performance, clean the condenser from time to time. Switch off the unit, disconnect mains power cable. Hold the venting grid, pull out and remove.