Owner manual

HL
89
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of circulator
version None Version number of the software (V X.xx)
status none Status message, error message (see page 91)
in_pv_00 none Current bath temperature.
in_pv_01 none Heating power being used (%).
in_pv_02 none Temperature value registered by the external Pt100 sensor.
in_pv_03 none Temperature value registered by the safety sensor.
in_pv_04 none Setpoint temperature (“SafeTemp”) of the excess
temperature protection
in_sp_00 none Working temperature “Setpoint 1”
in_sp_01 none Working temperature “Setpoint 2”
in_sp_02 none Working temperature “Setpoint 3”
in_sp_03 none High temperature warning limit “OverTemp”
in_sp_04 none Low temperature warning limit “SubTemp”
in_sp_05 none Setpoint temperature of the external programmer
(REG+E-PROG).
in_sp_06 none Temperature indication in °C or °F
in_sp_07 none 1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotational
speed after start.
in_sp_08 none Value of a flow rate sensor connected to the E-Prog input
in_sp_09 none Value of a pressure sensor connected to the E-Prog input
in_par_00 none Temperature difference between working sensor and safety
sensor
in_par_01 none Te - Time constant of the external bath.
in_par_02 none Si - Internal slope
in_par_03 none Ti - Time constant of the internal bath.
in_par_04 none CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05 none Factor pk/ph0:
Ratio of max. cooling capacity to max. heating capacity
in_par_06 none Xp control parameter of the internal controller.
in_par_07 none Tn control parameter of the internal controller.
in_par_08 none Tv control parameter of the internal controller.
in_par_09 none Xp control parameter of the cascade controller.
in_par_10 none Proportional share of the cascade controller.
in_par_11 none Tn control parameter of the cascade controller.