5. Preparations
5.1. Installation
Place the unit on an even surface on a pad made of non-flammable
material.
Pump motor and electronics produce intrinsic heat that is dissipated
via the venting openings! Never cover these openings!
Do not set up the unit in the immediate vicinity of heat sources and
do not expose to sun light
The bath cover (1) is mounted on the bath tank with 4 captive screws (2).
Required tool: SW3 hexagonal socket wrench with ball-shaped head
The circulating pump transports the
bath fluid to the temperature
chamber (4) at the bottom (3). The
overflow on top of the chamber
ensures a constant fluid level during
operation.
The bath cover (5) provides 6
measuring places (6) that can be
fitted with different Viton
®
sleeves..
Order-No.: Viton
®
sleeves
8 930 602 2 mm inner dia.
8 930 603 3 mm inner dia.
8 930 604 4 mm inner dia.
8 930 605 5 mm inner dia.
8 930 605 6 mm inner dia.
8 930 608 8 mm inner dia.
Take off the complete insert for cleaning. The gasket below the bath
cover should be replaced in this case.
Part No.: 7.901.1087 Gasket for bath cover 8K / 14K
5.2. Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of 65 °C is used, only supervised operation is possible.
Water: The quality of water depends on local conditions.
Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to calcification in the bath.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.