6.3. Start - Stop
Start:
Press
key.
The actual bath temperature is displayed on the VFD COMFORT-
DISPLAY. The circulating pump starts with a slight delay.
Stop:
Press
key.
or
Keep
key pressed.
The VFD COMFORT-DISPLAY indicates the message "OFF".
6.4. Adjusting the temperature chamber
Adjusting the temperature chamber after set
up:
If bath fluid overflow in the temperature
chamber is not equal in every direction, an
adjustment can be made with the 3 screws
that are accessible from above.
Required tool: SW3 hexagonal socket wrench
with ball-shaped head