Manual

Remote control
82
Command Parameter Response of circulator
out_mode_03 0 Set external programmer input to voltage.
Voltage 0 V ... 10 V
out_mode_03 1 Set external programmer input to current.
Current 0 mA ... 20 mA
out_mode_04 0 Temperature control of internal bath.
out_mode_04 1 External control with Pt100 sensor.
out_mode_05 0 Stop the unit = R –OFF-.
out_mode_05 1 Start the unit.
out_mode_08 0 Set the control dynamics - aperiodic
out_mode_08 1 Set the control dynamics - standard
out_sp_00 xxx.xx Set working temperature. “Setpoint 1”
out_sp_01 xxx.xx Set working temperature. “Setpoint 2”
out_sp_02 xxx.xx Set working temperature. “Setpoint 3”
out_sp_03 xxx.xx Set high temperature warning limit “OverTemp”
out_sp_04 xxx.xx Set low temperature warning limit “SubTemp”
out_sp_06 xxx.xx Set manipulated variable for heater via serial interface
-100 ... +100 [%]
out_sp_07 x Set the pump pressure stage. (1 ... 4)
out_par_04 x.x CoSpeed for external control 0 ... 5.0.
out_par_06 xxx Xp control parameter of the internal controller. 0.1 ... 99.9
out_par_07 xxx Tn control parameter of the internal controller. 3 ... 9999
out_par_08 xxx Tv control parameter of the internal controller. 0 ... 999
out_par_09 xxx Xp control parameter of the cascade controller. 0.1 ... 99.9
out_par_10 xxx Proportional share of the cascade controller. 1 ... 99.9
out_par_11 xxx Tn control parameter of the cascade controller. 3 ... 9999
out_par_12 xxx Tv control parameter of the cascade controller. 0 ... 999
out_par_13 xxx Maximum internal temperature of the cascade controller.
out_par_14 xxx Minimum internal temperature of the cascade controller.
out_par_15 xxx Band limit (upper) 0 ... 200 °C
out_par_16 xxx Band limit (lower) 0 ... 200 °C
out_hil_00 -xxx Set the desired maximum cooling power (0% to 100%).
Note: Enter the value with a preceding negative sign!
This setting is meaningful only for FP cooling machines.
out_hil_01 xxx Set the desired maximum heating power (10% to 100%).