User Manual

39
8 Configuration
8.3 Generator “Pro”
The basic function of the instrument is defined here. The instrument can be operated
as a fixed-setpoint controller with or without a ramp function, or warm-up ramp for
hot-channel equipment, program controller or program generator.
Configuration
Analog inputs
Controller
Generator
Limit comparators
Outputs
Binary functions
Display
Timer
Interfaces
Symbol Value/selection Description
General
Function
Fnct
0
1
2
3
4
Fixed-setpoint controller
Ramp function
Program controller
Program generator
Hot-channel controller
Ramp function:
A rising or a falling ramp function can be implemented. The
ramp end value is determined by the setpoint input.
The ramp function can be paused or canceled via the binary
functions.
v Chapter 8.6 “Binary functions “binF””
Program generator:
The setpoint profile is output via a continuous output.
Unit of slope
Unit
0
1
2
Ramp function Program
°C/min mm:ss
°C/hour hh:mm
°C/day dd:hh
s=seconds; m=minutes; h=hours;d=days
Unit of ramp slope in °C per time unit, or format of segment
times for program controller/generator.
Factory settings are shown bold.
H
The ramp function is interrupted on a probe break, or
for manual mode. The outputs react as for overrange/
underrange (configurable).