Technical data
General
Input
Reference conditions DIN 16086 and EN 60770
Sensor system Silicon sensor with stainless steel separating diaphragm
Pressure transfer medium Synthetic oil (silicon oil), FDA-compliant oil
Permissible load changes > 10 million
Location
Mounting position Any
Calibration position Device standing upright, process connection on bottom
Position-dependent zero point offset Zero point correction possible locally or via setup (20% of nominal measuring range).
- Basic type 000 standard ≤ 1 mbar
- Basic type 004 high-temperature design ≤ 10 mbar
Display Positively lit display
Alignment Display can be rotated 180° via the software
Enclosure rotatable ±160° (use the combination tool supplied)
Size Display field 16 x 26 mm / font size 7 mm / 2 x 4-digit
Color Normal operation: amber-colored
Error: red (text “Err”, error code 1 - 9 flashes)
Setup interface busy: red
Switching state indicator K1, K2
Measuring unit mbar, bar, kPa, MPa, psi, %
Operation
Local Via control element under the screw plug with the combination tool or 0.5x3 screwdriver or 2AF
hex key
Setup interface Pin 5 of M12x1 round plug
Relative pressure Nominal measuring ranges start at 0 bar.
Nominal measuring range 0.4 1 4 10 25 60 bar
Overload capacity
a
a
All measuring ranges can be overloaded to -1 bar (vacuum-proof).
1.6 4 16 40 100 240 bar
Bursting pressure 2 5 20 50 125 300 bar
Nominal measuring range -0.4 to 0.4 -1 to 1 -1 to 3 -1 to 9 -1 to 24 bar
Overload capacity
a
1.6 4 16 40 100 bar
Bursting pressure 2 5 20 50 125 bar
Absolute pressure Nominal measuring ranges start at 0 bar.
Nominal measuring range 0.4 1 4 10 25 60 bar
Overload capacity
a
1.6 4 16 40 100 240 bar
Bursting pressure 2 5 20 50 125 300 bar