User Manual
11 Instrument description
50
Input
All measuring ranges can be overloaded to -1 bar (vacuum-proof)
Outputs
Relative pressure
Measuring range 0.4 1 4 10 25 60
barOverload capacity 1.6 4 16 40 100 240
Bursting pressure 2 5 20 50 125 300
Measuring range -0.4...0.4 -1...1 -1...3 -1...9 -1...24
barOverload capacity 1.6 4 16 40 100
Bursting pressure252050125
Absolute pressure
Measuring range 0.4 1 4 10 25 60
barOverload capacity 1.6 4 16 40 100 240
Bursting pressure 2 5 20 50 125 300
Analog output
Freely configurable
Jump response T
90
4 ... 20 mA + 1 x PNP switching output standard
0 ... 20 mA + 1 x PNP switching output
0 ... 10 V + 1 x PNP switching output
≤ 100 ms
Switching output
Number
Switching type
Switching function
1 x PNP switching output
2 x PNP switching output
Break contact / make contact
Window / hysteresis
Switching capacity
- Voltage drop from U
B
- Contact rating
- Switching cycles
Response time
Short-circuit proof
PNP ≤ 2V
On ≤ 250 mA / Off ≤ 1mA
> 10 million
≤ 20 ms
Yes
Current load check
- Time interval
- Periodic protective
circuit in case of
overcurrent
2s; T
ON
40 ms
f=0.5 Hz
Display: Err3 switching output K1, Err4 switching output K2
Setting range
Analog output Turn down 1:4