User Manual

53
11 Instrument description
Accuracy
Relative pressure
Nominal (rated)
measuring range
0.414102560
bar
Linearity
1
0.15 0.15 0.1 0.1 0.1 0.1 % of end val. of
nom. mess. range
Accuracy at
+20°C
2
0.35 0.3 0.25 0.25 0.25 0.25 % of end val. of
nom. mess. range
Accuracy at
-20...+75°C
3
0.7 0.6 0.5 0.5 0.5 0.5 % of end val. of
nom. mess. range
Nominal (rated)
measuring range
-0.4...0.4
-1...1 -1...3 -1...9 -1...24 bar
Linearity
1
0.15 0.15 0.1 0.1 0.1 % of end val. of
nom. mess. range
Accuracy at
+20°C
2
0.35 0.3 0.25 0.25 0.25 % of end val. of
nom. mess. range
Accuracy at
-20...+75°C
3
0.7 0.6 0.5 0.5 0.5 % of end val. of
nom. mess. range
Absolute pressure
Nominal (rated)
measuring range
0.41 4 102560bar
Linearity
1
0.15 0.15 0.1 0.1 0.1 0.1 % of end val. of
nom. mess. range
Accuracy at
+20°C
2
0.35 0.3 0.25 0.25 0.25 0.25 % of end val. of
nom. mess. range
Accuracy at
-20...+75°C
3
0.7 0.6 0.5 0.5 0.5 0.5 % of end val. of
nom. mess. range
Long-term stability 0.2% with reference conditions to EN 61298-1
1
Linearity based on limit point setting
2
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of
measuring range.
3
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of
measuring range, thermal effect on initial value of measuring range and span.