Power supply
Mechanical properties
Burden
for output 405 (4 to 20 mA) Burden (U
B
-11,5 V)/0,022 A
for output 410 (4 to 20 mA with HART
®
)Burden (U
B
-11,5 V)/0,022 A; additional: min. 250 , max. 1100
For version
Explosion protection 0 (none) 11,5 to 36 V DC
Explosion protection 1 (ATEX Ex ia) 11,5 to 28 V DC
The power supply must be intrinsically safe
and must not exceed the following maximum values:
U
i
28 V DC
I
i
93 mA
P
i
750 mW
Process connection
Material Stainless steel 316L
for nominal measuring range 516 Stainless steel 1.4542
Surface Ra 0.8 µm
Process seal
for process connection 512 and 571 FPM
for process connection 652 FPM
for process connection 997
(JUMO PEKA)
FDA-compliant, EHEDG approved materials: FPM, VMQ, EPDM option, see data sheet 409711
for all other process connections Without a seal
Messmembrane
Material Stainless steel 316L
Surface Ra 0.8 µm
Housing
Material
for housing 1 (short, stainless steel) Stainless steel 1.4404
for housing 2 (long, stainless steel) Stainless steel 1.4404, VMQ
for housing 3 (precision casting) Precision casting 1.4408
for cover material 20 (stainless steel) Precision casting 1.4408, FPM seal
for cover material 85 (plastic) PA, FPM seal
for electrical connection 36
(M12 × 1 round plug)
Nickel-plated brass
for electrical connection 82
(cable gland, plastic)
PA
for electrical connection 93
(cable gland, metal)
Nickel-plated brass
for operation 0 (without control knob) -
for operation 1 (with control knob) PA
Explosion protection
for explosion protection 0 (none) The instrument is not approved for use in hazardous (Ex) areas.
for explosion protection 1 (ATEX Ex ia) EC type examination certificate SEV 09 ATEX 0138 X
II 1/2G Ex ia IIC T6 to T3 Ga/Gb
II 1/2D Ex ia IIIC T105 °C Da/Db
Weight
type 403025/0-0-1 (short housing) approx. 550 g
type 403025/0-0-2 (long housing) approx. 850 g
type 4030252/0-0-3
(precision casting housing)
approx. 1600 g