Datasheet

2013-03-12/00535026
Data Sheet 403025 Page 4/11
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax: 315-437-5860
E-mail: info.us@jumo.net
Internet: www.jumousa.com
Ambient conditions
Accuracy
Permissible temperatures
Operation Version Category Max. medium
temperature
Environment
temperature
a
a
Restricted function below -20 °C: stationary use, increased danger of broken cable, display does not function.
Extended environment
temperature
(extra code 681)
a, b
b
In the range of -40 to -50 °C, the cover with the instrument viewing pane must also be protected against mechanical shock and impact.
For details please contact JUMO.
Standard +120 °C -40 to +85 °C -50 to +85 °C
High temperature +200 °C -40 to +85 °C -50 to +85 °C
II 1/2G Ex ia T6 +60 °C -40 to +50 °C -50 to +50 °C
T5 +70 °C -40 to +65 °C -50 to +65 °C
T4 +115 °C -40 to +85 °C -50 to +85 °C
T3
c
c
Only for high temperature version (temperature of medium 2).
+175 °C -40 to +85 °C -50 to +85 °C
II 1/2D Ex ia T105 °C +100 °C -40 to +60 °C -50 to +60 °C
Storage -40 to +85°C
Permissible relative humidity
Operation 100 % incl. condensation on instrument outer sleeve
Storage 90 % without condensation
Permissible mechanical loading
Vibration performance 2 g, 10 to 2000 Hz to DIN EN 60068-2-6
Shock resistance 15 g for 6 ms to IEC 60068-2-27
Electromagnetic compatibility To EN 61326
Interference emission Class B
Interference resistance Industry
Protection
Version
Explosion protection 0 (none) IP67 to DIN EN 60529
Explosion protection 1 (ATEX Ex ia) IP66 to DIN EN 60529
Relative pressure
Nominal measuring range -600 to 600 mbar
rel.
-1 to 4 bar rel. -1 to 25 bar rel. -1 to 100 bar rel. -1 to 600 bar
rel.
Factory setting for measuring range 0 to 600 mbar 0 to 4 bar 0 to 25 bar 0 to 100 bar 0 to 600 bar
Minimum span 60 mbar 0.1 bar 0.5 bar 5 bar 30 bar
Turndown ratio (r) r 20 r 50 r 52 r 20
Linearity for linear characteristic
as % of the set span
0.1 % for r 10 0.05 % for r 10
r × 0.01 %
for 10 r 20
r × 0.005 %
for 10 r 50
r × 0.005 %
for 10 r 52
r × 0.005 %
for 10 r 20
Accuracy at 20 °C
as % of the set span
0.2 % for r 10 0.1 % for r 10
r × 0.02 %
for 10 r 20
r × 0.01 %
for 10 r 50
r × 0.01 %
for 10 r 52
r × 0.01 %
for 10 r 20
Accuracy at -40 to +85 °C
as % of the set span
0.3 % for r 10
a
a
Only to -30 °C.
0.2 % for r 10 0.2 % for r 10
r × 0.03 %
for 10 r 20
a
r × 0.02 %
for 10 r 50
r × 0.02 %
for 10 r 52
r × 0.02 %
for 10 r 20
Long-term stability
as % of nominal measuring range
0.1 %/year