JUMO BlackLine CR-GT
with the cell constants K = 1.0
The GT version features a special graphite as the electrode material. The graphite has been
especially treated to achieve - in conjunction with suitable evaluation devices - measuring
ranges up to 100 mS/cm and above.
A temperature probe can optionally be integrated. The conductivity sensor with 120 mm fitting
length and a shaft diameter of 12 mm can be used in combination with standard pH or redox
electrodes in suitable built-in fittings. The sensor is also ideally suited for linking with handheld
or laboratory measuring equipment. When used as an online sensor, suitable flow-through
fittings must be employed (see JUMO data sheets 202810, 202820, 202822).
JUMO BlackLine CR-EC
with the cell constants K = 0.01, K = 0.1 and K = 1.0
The principal features of the EC version are its very compact construction and the Pt100 that is
integrated as standard. With a fitting length of just 40 mm (or 63 mm), the sensors can also be
screwed into pipelines with a small nominal diameter. To this end, the conductivity sensors
either have a G 1/2A or 1/2"-14 thread.
The version with K = 1.0 achieves a measuring range of 0.1 to about 5000 µS/cm (5mS/cm).
Titanium pins function as the measuring electrodes here. The version with K=1.0 has been
implemented as a coaxial cell and can be used to about 1000 µS/cm. Sensors with K = 0.01
cover ranges from 0.05 to 20 µS/cm. They are therefore extremely suitable for pure and high-
purity water applications. The electrode material for these sensors is stainless steel 1.4571.
JUMO BlackLine CR-GS glass conductivity sensors K = 1.0
Glass conductivity sensors are used to conductively determine the electrolytic conductivity of
aqueous solutions. The cell constant of the sensor is K = 1.0
1
/
cm
. These sensor can be supplied
with an integrated Pt100 masuring insert, as an option.
Versions with a Pg13.5 threaded cap can be incorporated in suitable fittings (sse data sheets
202810, 202820, 20,2822 and 202825).
Operating principle
The sensors in the 202922 series are 2-electrode conductivity sensors. A transmitter applies an
AC voltage to the sensors. The current flowing through the liquid and the electrodes is
determined by the conductivity of the liquid.
Data Sheet 202922
V2.01/EN/00437028
Vorläufig
V
JUMO BlackLine CR-GT/-EC/-GS
Conductive 2-Electrode Conductivity Sensor
Cell constants K = 0.01 – 1.0
202922 Series
for universal application, e.g. drinking and surface water monitoring, ion exchangers and
reverse osmosis plant, construction of air conditioning and cooling systems, horticultural
technology, sea water and fresh water aquaculture, mildly contaminated industrial rinse and
process water, high-purity water monitoring, etc.
compact style for universal application
manufactured from physiologically safe materials
JUMO BlackLine CR-GT (left),
JUMO BlackLine CR-EC (middle),
JUMO BlackLine CR-GS (right)
20292200T10Z001K000