7 Technical data
Analog inputs (channels 1 — 12)
Thermocouple
Resistance thermometer
Designation Type Standard Range Linearization accuracy
1
Fe-Con L DIN 43 710
Fe-Con J EN 60 584
Cu-Con U DIN 43 710
Cu-Con T EN 60 584
NiCr-Ni K EN 60 584
NiCr-Con E EN 60 584
NiCrSi-NiSi N EN 60 584
Pt10Rh-Pt S EN 60 584
Pt13Rh-Pt R EN 60 584
Pt30Rh-Pt6Rh B EN 60 584
Chromel-Copel
-200 to +900°C
-210 to +1200°C
-200 to +600°C
-270 to +400°C
-270 to +1372°C
-270 to +1000°C
-270 to +1300°C
-50 to +1768°C
-50 to +1768°C
0 to 1820°C
-200 to +800°C
±0.1%
±0.1% above -100°C
±0.1% above -150°C
±0.15% above -150°C
±0.1% above -80°C
±0.1% above -80°C
±0.1% above -80°C
±0.15% above 0°C
±0.15% above 0°C
±0.15% above 400°C
±0.15%
Shortest span L, J, U, T, K, E, N, Chromel-Copel:
S, R, B:
100°C
500°C
Range start/end freely programmable within the limits in 0.1°C steps
Cold junction Pt100 internal or thermostat external constant
Cold junction accuracy (internal) ± 1°C
Cold junction temperature
(external)
-50 to +100°C adjustable through setup software
Sampling cycle 6 or 12 channels 125msec
Input filter 2nd order digital filter; filter constant adjustable from 0 to 10.0sec
Test voltage 500V (across optocoupler)
Resolution >14 bit
Features also programmable in °F
1.The linearization accuracy refers to the maximum span.
The linearization accuracy is reduced for shorter spans.
Designation Standard Connection Range Linearization
accuracy
Measuring
current
Pt 100 EN 60 751 2/3-wire
2/3-wire
4-wire
4-wire
-200 to +500°C
-200 to +850°C
-200 to +500°C
-200 to +850°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
500μA
250μA
500μA
250μA
Pt 100 JIS 2/3-wire
2/3-wire
4-wire
4-wire
-200 to +500°C
-200 to +650°C
-200 to +500°C
-200 to +650°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
500μA
250μA
500μA
250μA
Pt 500 EN 60 751 2/3-wire
2/3-wire
4-wire
4-wire
-200 to +500°C
-200 to +850°C
-200 to +500°C
-200 to +850°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
250μA
250μA
250μA
250μA
Pt 1000 EN 60 751 2/3-wire
2/3-wire
4-wire
4-wire
-200 to +500°C
-200 to +850°C
-200 to +500°C
-200 to +850°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
500μA
250μA
500μA
250μA
Ni 100 EN 43760 2/3-wire
4-wire
-60 to +180°C
-60 to +180°C
±0.4°C
±0.4°C
500μA
500μA