Instruction Manual

9
2) The fetal movement pen will automatically turn on when it is removed from the side of the host.
3) When used, pregnant women press the fetal movement pen button when they feel fetal movements or
uterine contractions, the host will record the fetal movement. It records once only if it is repeatedly pressed within 15
seconds.
3.2.4. Shuttle button
1) Adjust the volume: in monitoring mode, select the shuttle button to adjust the volume. There are eight
grades totally from 0 to 7. The corresponding system volume is from silent to the strongest, which can be
adjusted by shuttle button, reducing sound through rotating to the left and increasing sound through rotating
to the right.
2) Freeze playback: in monitoring mode, short press the shuttle button to enter the freeze playback function.
The monitoring information can be played back by rotating the shuttle button (fetal heart sound does not
support playback).
3) Selection and confirmation: in setting mode, rotating shuttle button is the selection function. The cursor will
move when shuttle button rotates. Short press the shuttle button to confirm if enter the submenu or select a
parameter.
3.2.5. Uterine contraction back-to-zero button
1) In monitoring mode, short press this button after the uterine contraction probe is fixed by bandage, the
pressure display value is reset to 10.
3.2.6. Awaking button
1) Press once to awake fetal heart probe once, the awaking icon is displayed once on the pressed
time
.
3.2.7. Ultrasonic fetal heart probe
1) When the host is in power on state, the ultrasonic fetal heart probe is taken out and turned on automatically.
The probe will show the current electric quantity and it will disappear after 3 seconds, then the signal light is on.
During the monitoring process, the signal light flashes with the signal strength of the ultrasonic fetal heart probe;
2) The ultrasonic fetal heart probe is put into the host, making the metal contact of the fetal heart probe touch
the sub-contact of the host charging terminal. The probe will automatically pair with the host, all power indicator
lights are on, indicating the pairing is successful. The host will automatically charge for the probe;