1. Normal work: display the current electric quantity;
2. Low electric quantity tip: when the electric quantity is too low, it will flash continuously;
3. Charging: it displays charging when access to the power adapter, the charging symbol flashes;
Bed No.: displays the current bed number. the bed number is consistent with the device ID.
Name: displays the name of the pregnant woman who is currently being monitored (connection to the central station
is required).
Duration: the monitoring duration is started from pressing the start key.
Fetal heart rate: displays the current fetal heart rate. The abnormal icon will light up if the current fetal heart rate
exceeds the upper and lower limits of 15 seconds.
Uterine contraction: displays the current uterine contraction value.
Fetal movement: press a fetal movement mark to calculate the fetal movements. It records once only if it is
repeatedly pressed within 15 seconds.
Fetal heart probe connection label: indicates the state label connected with the ultrasonic fetal heart probe. When
the host is paired with the ultrasonic fetal heart probe and the probe is on, the white background of the label will be on
and the current electric quantity of the fetal heart probe will be displayed. When the fetal heart probe is taken out from
the host box, the label will become green, indicating that automatic matching is complete. You can use the probe for
fetal heart monitoring.
Uterine contraction probe connection label: indicates the state label connected with the uterine contraction probe.
When the host is paired with the uterine contraction probe and the probe is on, the white background of the label will
be on and the current electric quantity of the uterine contraction probe will be displayed. When the uterine contraction
probe is taken out from the host box, the label will become green, indicating that automatic matching is complete. You
can use the probe for uterine contraction monitoring.
Fetal movement record: indicates that the user presses the fetal movement pen at that time.
Awaking record: indicates that the awaking function has been used once at that time.
3.4. Setting interface specification