Instruction Manual

22
Probe damaged
Change the probe
Abnormal uterine
contraction pressure
The probe is not connected to the
device well
Incorrect probe position
No pressure reset
Pregnant woman has no uterine
contraction
Reconnect
Readjust the probe position
Process pressure reset
Wait for pregnant woman’s uterine
contraction
Press the pressure probe,
the value of uterine
contraction has no change
or little change.
Circuit drift in pressure probe
Pressure probe damaged
Contact the manufacturer for
maintenance or Change the pressure
probe
No response when press
the FM pen
The FM pen is not successfully
matched with the host
Place the FM pen in the side of the
device for re-adsorption, and then take
out the FM pen again
Speaker silence
Too low volume adjustment
Volume up
WiFi icon shows gray
Turn off WiFi function
Disconnect from the router
Access to setting interface, open the
WiFi function and configure WiFi
Check the router
The screen displays the
probe icon in gray
Unpaired
The probe is charging
Retake the probe and align to the
contact point for matching.
Normal phenomenon
9. Warranty Period and Maintenance
We provide the one-year warranty and lifelong maintenance for this product. If the device fault occurs to device
fails to work, the manufacturer shall be contacted in time, any form of maintenance or other disposal of the
device is prohibited.
The service life of this device is five years.
Please contact the manufacturer for repair or replacement if any problem occurs in the following
accessories:
Probe: When the probe line appears to be damaged, broken or in bad contact, please contact the manufacturer in
time to consult for maintenance or replacement.
Power line: When the power line appears to be damaged, broken or in bad contact, please immediately stop to
use, and consult the manufacturer to undertake timely replacement.
About after-sales service, maintenance and any other involved problem or any problems involving the company's