Instruction Manual
Table Of Contents
- FCC Statement
- 1. Introduction
- 1.1. Product Introduction
- 1.2. Safety Guide
- 1.3. Symbol and text description
- 2. Recommended clinical application
- 3. Product Description
- 3.1. Standard Configuration
- 3.2. Product Introduction
- 3.3. Main interface (monitoring interface) descrip
- 3.4. Setting interface specification
- 4. Setting of monitor
- 4.1. Disassembly and installation and list of acce
- 4.2. Connect the power supply Charge
- 5. Operational Guidelines
- 5.1. Preparation before Use
- 5.2. FHR monitoring
- 5.3. TOCO monitoring
- 5.4. Manual FM monitoring
- 5.5. Monitoring information update
- 5.6. Restore Factory Settings
- 6. Maintenance and clean of Fetal Monitor
- 6.1. Product Handling
- 6.2. Maintenance
- 6.3. Cleaning and Disinfection
- 6.4. Gel
- 7. Packaging, transportation and storage
- 7.1. Packaging
- 7.2. Transport
- 7.3. Storage
- 8. Fault finding and eliminating
- 9. Warranty Period and Maintenance
- 10. Technical Parameters
- AppendixⅠ—Electromagnetic Compatibility Declaratio
24
Signal processing: special digital signal processing unit
Measuring range: 60bpm to 210bpm
Alarm Range:
Upper limit: the adjustable range is 150bpm to 180bpm, the default is 160bpm
Lower limit: the adjustable range is 90bpm to 120bpm, the default is 110bpm
Maximum audio output power: 2W
UC pressure
Measuring range: 0-100mmHg
Display
The LCD screen shows the monitoring state of FHR curve, UC pressure curve, FM symbols, volume and so on,
which can be frozen and replayed.
Power supply: AC 100-240V 50/60Hz, internal battery: lithium battery3.7V 3000mAh
Input Power: 20VA
Work mode: continuous working for more than 10 hours after full power supply
Size: 89.2mm (length)*202mm(width)*75.9mm (height)
Net weight: 882.5g (power adapter and accessories are not included)
Environment
Working environment: temperature: +5℃ to +40℃ humidity: ≤ 80% atmospheric pressure: 86KPa to 106KPa
Transportation and storage environment: temperature: -20 ℃ to +55 ℃ humidity: ≤ 95% atmospheric pressure:
50KPa to 106KPa
Probe sound output: According to the regulations of international standard IEC 61157, the peak negative acoustic
pressure shall not exceed 1 MPa, and the output beam sound intensity shall not exceed 20mW/cm
2
. The