Instruction Manual

28
environment
guidance
RF emissions
CISPR 11
Group 1
The JPD-300E uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely
to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The JPD-300E is suitable
for use in all
establishments, including
domestic establishments and
those directly connected to
the public low-voltage power
supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuation/ flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration electromagnetic immunity
The JPD-300Eis intended for use in the electromagnetic environment specified below. The customer or
the user of the JPD-300E should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance level
Electromagnetic
Environment -
Guidance
Electrostatic discharges
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8kV Contact
±15kV Air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
If ESD interfere with
the operation of
equipment,counter
measurements
such as wrist strap,
grounding shall be
considered.
Electrical fast transient
(EFT)
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality
should be that of a
typical commercial or