hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Voltage dips,short
Interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % UT
(100 % dip in UT)
for 0,5 cycle
0 % UT
(100 % dip in UT)
for 1 cycles
70 % UT
(30 % dip in UT)
for 25/30cycles
0 % UT
(100 % dip in UT)
for 250/300 cycles
0 % UT
(100 % dip in UT)
for 0,5 cycle
0 % UT
(100 % dip in UT)
for 1 cycles
70 % UT
(30 % dip in UT)
for 25/30cycles
0 % UT
(100 % dip in UT)
for 250/300 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
JPD-300E requires
continued operation
during power mains
interruptions, it is
recommended that the
JPD-300E be powered
from an uninterruptible
power supply or a
battery.
Power frequency
magnetic field
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels haracteristic
of a typical location
in a typical
commercial or hospital
environment.
Note: U
T
means AC network voltage before applied with test voltage.
Guidance and manufacturer s declaration electromagnetic immunity for EQUIPMENT and
SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration electromagnetic immunity
The JPD-300E is intended for use in the electromagnetic environment specified below. The customer or
the user of the JPD-300E should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Compliance
level
Electromagnetic Environment
- Guidance
Conducted
RF
IEC 61000-4-6
3 Vrms 150 kHz to
80 MHz 6V in ISM
and amateur radio
bands between
0.15MHz and 80MHz
3 Vrms 150
kHz to
80 MHz6V in
ISM and
amateur radio
bands between
Portable and mobile RF
communications equipment
should be used no closer to any
part of the JPD-300E, including
cables, than the recommended
separation distance calculated