MCD 8000 Performance Specifications
Reference Operating Conditions (ROC):
+/-10% line voltage
23+/-2
o
C ambient temperature
<80% RH non-condensing
Accuracy (at ROC): 0.02% of reading +/-1 count
Noise Rejection:
NMRR: >-60dB @ 50/60Hz, +/-.1Hz
CMRR: >-120dB @ 50/60Hz, +/-.1Hz with 250 ohm unbalanced
Overload Protection:
Power lead to ground: 1500VDC or ACRMS
Across inputs: Voltage up to 250VDC or VAC for 1 min., +V to -V
Current up to 150mADC or mAAC for 1 min., +I to -I
Stability with Temperature: Zero 1uV/
o
C, Span 0.01% reading/
o
C
Stability with Time: 10 counts per year maximum
Repeatability: +/- 1 count
MCD 8000 Environmental/Physical Specifications
Operating Range:
Temperature: 5 to 45
o
C
Relative Humidity: 10-80% RH non-condensing
Storage Range: -40 to 65
o
C
Power:
115VAC +/-10%, 50-60 Hz