Owner manual
28
6.5 To Access From Your PC
To connect IAQ Monitor to your PC, please
refer to p.27.
―――Icons and its Meaning―――
:
Space
: Return or Press Enter
*: A Number
※Please input all commands with a capital letter.
6.5.1 Transmission of On-Time data
Display Procedure
Example:
Measuring Model and typed Entered
[D0005
].
<Set Number of Data Needed>
Press “D****
” (※Must type in 4 digits)
After “AD”, the data will be displayed.
Each data represent 1sec of measurement. If you ask for 20 data, it
takes approximately 20sec to display.
The maximum number that can be set is 999. If you need more,
re-send the command.
<Output Content>
CO; CO2; Temperature; Humidity
<To Cancel>
Press “N
”
Display Procedure
<To Download Measuring Conditions>
Press
[S ]
After “AS”, the data will be displayed.
Contents
CTH;CO range; temperature range; atmosphere pressure
CO range -- 00: 0-50PPM, 01: 0-500PPM
Temperature range – 00: 0-60℃
, 01: -20-60
℃
AN
AD
0.9; 576; 23.4; 63.4
0.8; 556; 23.4; 63.3
0.8; 534; 23.5; 63.2
0.9; 540; 23.5; 63.2
0.9; 561; 23.4; 63.3
AS
CTH;00;00;1013
Command Function
D**** Number of Downloading Data
N Cancel
S Output of Measuring Condition (of
On-Time Data)
U Output of Measuring Units
K Output of Duct Shape/Size
P Output of Page Number
T**** Output of Stored Data
M**** Output of Measuring Condition (of
Stored Data)
B Output of Measuring Condition of
All Pages