Display Procedure
<To download measurement condition>
Press M****
(Must type in 4 digits)
Type in the desired page number after AM.
Contents
Measuring Mode
000: CALCULATION Mode
001: %OA Mode
Sampling Time
Number of Data
Calculation setting
CALCULATION -- AVG: average, INS: Instant
%OA -- TMP: temperature CO2; CO
2
Atmosphere pressure
Display Procedure
<To download measurement conditions of all pages>
Press
[B ]
, after[AB]
The data will be displayed.
Display Procedure
<Error Message>
"ED" will be returned if the number of pages etc. is incorrect-inputted.
6.6 Analog Output (Optional)
Data Update Interval ………1 second
Load Impedance…… Above 5KΩ
Output Current………………DC 0-1V
For the analog output, you must select one setting from the table below.
Output Range Conversion Formula
(Voltage: )
0 – 50 ppm C = 50 x V ppm
0 – 100 ppm
C = 100 x ppm
0 – 250 ppm
C = 250 x
ppm
CO C
0 – 500 ppm
C = 500 x ppm
0 – 500 ppm
M = 500 x
ppm
0 – 1000 ppm
M = 1000 x ppm
0 – 2500 ppm
M = 2500 x
ppm
CO
2
M
0 – 5000 ppm
M = 5000 x ppm
0 – 50
T = 50 x
0 –100 T = 100 x
-20 – 30
T = 50 x
20
Temperatur
eT
-20 – 80 T = 100 x 20
32 – 122
o
F
F = 90 x
+ 32
o
F
32 – 212
o
F
F = 180 x + 32
o
F
-4 – 86
o
F
F = 90 x
4
o
F
Temperatur
e(F)
-4 – 176
o
F
F = 180 x 4
o
F
0 – 50 %RH
H = 50 x %RH
Humidity
(H)
0 – 100 %RH
H = 100 x %RH
Analog Output Port
AM
CTH;000;001;003;AVG;1013
AB
CTH;000;001;003;AVG;1013
CTH;001;001;005;TMP;1013
ED