Owner manual
30
Display Procedure
<To download measurement condition>
Press “M****
” (※Must type in 4 digits)
Type in the desired page number after “AM”.
Contents
① Measuring Mode
000: CALCULATION Mode
001: %OA Mode
② Sampling Time
③ Number of Data
④ Calculation setting
CALCULATION -- AVG: average, INS: Instant
%OA -- TMP: temperature CO2; CO
2
⑤
Atmosphere pressure
Display Procedure
<To download measurement conditions of all pages>
Press
[B ]
, after[AB]
The data will be displayed.
Display Procedure
<Error Message>
"ED" will be returned if the number of pages etc. is incorrect-inputted.
6.6 Analog Output (Optional)
① Data Update Interval ………1 second
② Load Impedance…… Above 5KΩ
④ Output Current………………DC 0-1V
For the analog output, you must select one setting from the table below.
Output Range Conversion Formula
(Voltage: V)
0 – 50 ppm C = 50 x V ppm
0 – 100 ppm
C = 100 x V ppm
0 – 250 ppm
C = 250 x
V
ppm
CO (C)
0 – 500 ppm
C = 500 x V ppm
0 – 500 ppm
M = 500 x
V
ppm
0 – 1000 ppm
M = 1000 x V ppm
0 – 2500 ppm
M = 2500 x
V
ppm
CO
2
(
M
)
0 – 5000 ppm
M = 5000 x V ppm
0 – 50
℃
T = 50 x
V
℃
0 –100 ℃ T = 100 x V ℃
-20 – 30
℃
T = 50 x
V
– 20
℃
Temperatur
e(T)
-20 – 80 ℃ T = 100 x V – 20 ℃
32 – 122
o
F
F = 90 x
V
+ 32
o
F
32 – 212
o
F
F = 180 x V + 32
o
F
-4 – 86
o
F
F = 90 x
V
– 4
o
F
Temperatur
e(F)
-4 – 176
o
F
F = 180 x V – 4
o
F
0 – 50 %RH
H = 50 x V %RH
Humidity
(H)
0 – 100 %RH
H = 100 x V %RH
Analog Output Port
③
⑤
AM
CTH;000;001;003;AVG;1013
①
②
④
AB
CTH;000;001;003;AVG;1013
CTH;001;001;005;TMP;1013
ED