Owner manual

44
9. Specification
Product IAQ Monitor
Model 2211
Object Clean air flow
Method
Electrochemical
Range 0.0-50.0 ppm0.0~500 ppm*
Resolution 0.0-99.9ppm:0.1ppm,100-500ppm:1ppm
Accuracy
±3% of displayed value or ±3ppm; whichever that is larger @20℃)
Compensation: Temp
±0.125 %FS/ within 20-40: standard is 20℃)
Compensation:
Air Pressure
±0.02 %FS/hPa within 700-1200hPa: standard is 1013hPa
CO
Response Time
Approx. 60sec.
90%responsive, with calibration cover
Method
NDIR (Non-Distributed Infrared)
Range 0~5000 ppm
Resolution
1ppm
Accuracy
±3% of displayed value or ±50ppm; whichever that is larger @20℃)
Compensation: Temp ±
0.34%FS/
within
20-40
: standard is 20
℃)
Compensation:
Air Pressure
±0.02%FS/hPa within 700~1200hPa: standard is 1013hPa
CO2
Response Time
Approx. 45sec. 90%responsive, calibration cover
Method Platinum Temperature Measuring Resistive Element
Range
0.0~60.0°C(‐20.0~60.0
Resolution
0.1 °C
Accuracy ±0.5 °C
Temp.
Response Time Less than 60sec. velocity:m/s, 90% responsive
Method
Electrostatic Capacity
Range
2.0~98.0 %RH
Resolution 0.1 %RH
Accuracy
2~80%RH:±2.0%RH,80~98%RH:±3.0%RH
Humidity
Response Time Approx. 45sec.90%responsive
Function
Measured/Max value HOLD, Battery Indicator (5-level), Time/date,
Barometer comp., Unit selection Temp/DT/WB: °C or °F, AH:g/m
3
or
lb/ft
3
, OA%:g/kg or lb/lb, Max/Min/Average (interval: 1~999sec., retry:
1~999, Max memory: 1500data, OA%, Gas calibration
Output Function
Digital output: RS232C baud rate:4800,9600,19200,38400bps for
printer and/or PC
Analog Output: DC0~1V CO, CO2, temperature, or humidity
Power Supply 6x AA Batteries, AC Adaptor
*
: AC100~240V50/60Hz
Battery Life
Approx. 10hours
20, alkaline batteries, no RS232C connection
Main Body 5~40 no condensation
Probe 20~60 no condensation
Condition
Storage 20~60 no condensation
Weight Main body: Approx. 400g including batteries Probe: Approx. 250g
Standard Kits
Carrying case, Operation manual, AA batteries x6, Calibration cover and
tube, Probe stand, Data processing software for Windows, RS232C
cable
Optional Accesories
Spare probe, analog output, printer, ZERO gas, SPAN gas for CO/CO2,
flow control valve, AC adaptor