Manual

17
5.3 SINGLE Mode
By setting the sampling time, this unit automatically stops after the specified time.
The setting of the particle size (um), data storage (STR), alarm (BEEP) and printout (PR) are possible.
Display screen Operation key Operation explanation
POWER Push POWER key to input the power supply.
PREV Push PREV key twice to proceed MODE screen
SET
Select 2. SINGLE
SET
Setup the particle size, requirement of data
storage , alarm , printout, sampling time.
Use key to change the setting
condition, then push SET key.
After the input is done, change NO to OK
and push SET key.
WAIT Mode
WAIT sign is shown for the stabilization of
internal pump.
READY Mode
WAIT sign turns to READY in 10 seconds and
measurement can be started.
Push START/STOP key to start the
measurement
STARTSTOP
The display shows the real-time data.
Remaining time of each measurements
⑥:measurement number
INTERVAL menu
Screen changes to interval menu after the
sampling time are over.
The data is printed after the sampling time is
over if you chose printout. (refer to 6.4)
To stop the measurement halfway
Push START/STOP key.
Only the data of finished previous
measurement before pushing the STOP
key will be stored if you chose data storage
<MODE>
1. REPEAT 4.CALC
2. SINGLE 5.REMOTE
3. CONT
SINGLE 0.30.5µm
STR:Y BEEP:N PR:N
SAMPLE 10:00
OK
SINGLE 003BPM 15:25
WAIT 0.3 0.00E+0cf
0.5 0.00E+0cf
SINGLE 003BPM 15:25
READY 0.3 0.00E+0cf
0.5 0.00E+0cf
SINGLE 004BPM 15:25
09:59 0.3 0.00E+0cf
0.5 0.00E+0cf
SINGLE 004BPM 15:35
STOP 0.3 0.00E+0cf
0.5 0.00E+0cf
The particle size can be switched with the
key on the measurement screen.
UNIT: [CNT] [/m
3
] [/cf]