Instruction Manual

4
3. Precautions for Use
The following precautions must be taken when using the instrument.
Sampling
There are possibilities of particle deposit/rescattering when sampling is performed by using a
tube connected to the inlet. It is recommended that sampling is performed without using the tube.
However, if a tube is required for sampling, the tube recommend below must be used.
Sampling Tube
Recommended Sampling Tube:
TYGON Inner Diameter 4.3mm × Outer Diameter 7.5mm: Product or Norton
This product can be purchased from dealers of physical and chemical equipment, as well as
through our sales offices.
Power
The instrument can be powered by AA batteries or AC power.
AC Power
For AC-powered operation, the dedicated AC adapter supplied with the instrument must be used.
The AC adapter accommodates voltage of AC 86-264V 50/60Hz, however, the connection plug is
dedicated to AC 100V.
AA Batteries
In addition to the supplied AA rechargeable batteries (Ni-MH 1.2V 1600mA), alkali batteries can
be used as well.
Maximum continuous operating hours:
- When using alkali batteries : 1 hour
- When using the supplied rechargeable batteries : 3 hours
The above continuous operating hours are subject to change due to operational conditions such as
ambient condition, or use of the sampling tube.
Caution for Extremely Long Measurement Periods
The instrument is unsuitable for continuous measurements for a prolonged period. It may hasten
deterioration of the light source and inner pump, and require maintenance in a shorter interval.
For Prolonged Measurements
Note that the data will not be saved if the power is cut during a measurement.
Make sure to prevent running out of batteries or power cut of AC power.
Sampling Environment
The instrument is designed for use in clean rooms or clean environments where the concentration
is below the maximum measurable concentration of 2,000,000 particles/cf. Using the instrument
in a high temperature/humidity environment or in an environment with high particle
concentration may cause damage to the instrument or shorten the maintenance interval.