10. Troubleshooting
If you have a problem with your unit, please check the following list for solutions.
Symptom Possible Cause / Solution
Refer to
The display does not appear
when the power is turned ON.
The AC adapter is not connected properly.
Confirm the AC adapter and power cable.
Low battery
Replace the batteries.
Recharge the batteries (Ni-MH batteries)
3.1
Ni-MH battery drains fast. Insufficient battery charge Recharge 3.1
Reading is blinking. Alarm level is exceeded Confirm alarm setting 4.3
Measurement does not start.
When operating status is:
WAIT Wait until status changes to “READY”, and press
“START” key.
READY Press “START” key.
STOP Press “START” key, wait until status changes to
“READY”, and press “START” key again.
4
Particle count or concentration
is too high.
Attach the supplied filter and confirm that the reading drops
to zero.
Reading drops to zero:
Concentration of measuring environment is too high.
Reading remains high:
Possible instrument failure.
Please contact your local distributor.
Particle count or concentration
is too low.
Confirm error status of laser power failure or flow error. 8
Reading is displayed as “##.#” “##.#” indicates that the measurable range is exceeded.
Printing cannot be performed.
- Setting such as the baud rate setting is not made properly.
Confirm the printer setting.
- Improper cable connection.
(Confirm that the printer cable is used. Not the RS232C
cable.)
6.4
Can not transfer data to the
computer.
- Confirm cable connection.
(Confirm that the RS232C cable is used. Not the printer
cable. )
- Computer is not properly set for data transfer.
6.3
If the problem cannot be solved by confirming the above, please contact your local distributor or
our service center.