User's Manual
Table Of Contents
- Appendix B. Regulatory Compliance and Agency Approvals
- Chapter 1. Kaye ValProbe RT System
- Chapter 2. Kaye ValProbe RT Hardware
- Chapter 3. ValProbe RT System Connections
- Chapter 4. Creating User Accounts
- Chapter 5. Defining Equipment
- Chapter 6. Defining Assets
- Chapter 7. Defining Study Setups
- Chapter 8. Viewing Live Data
- Chapter 9. Qualification Study
- Chapter 10. Sensor Verification
- Chapter 11. Qualification Reports
- Appendix A. Specifications
- A.1 Kaye ValProbe RT System
- A.1.1 List of Certified Countries
- A.1.2 Regulatory Compliance
- A.1.3 RF System Operating Frequency and Modulation
- A.1.4 RF Channels
- A.1.5 RF Network Topology and PAN ID (Network ID)
- A.1.6 Antenna Diversity
- A.1.7 Data Rate and Data Encryption
- A.1.8 RF Range (Line of Sight and Outdoor)
- A.1.9 Number of Loggers and Sensors Supported
- A.1.10 Sensor Sampling Rate
- A.1.11 Data Transmission Rate
- A.1.12 Sensor Samples Data Storage Capacity
- A.1.13 Optional Extendable Base Station Antenna
- A.2 Kaye ValProbe RT Base Station
- A.2.1 Input Power and Adaptor Ratings
- A.2.2 Installation Category
- A.2.3 Overvoltage Category and Electrical Safety Protection Class
- A.2.4 Input DC Power On/Off Switch
- A.2.5 Front Panel LED Indications and Buzzer
- A.2.6 Bath and IRTD Interface Ports
- A.2.7 USB Port
- A.2.8 Ethernet Communication Port
- A.2.9 Reset Switch to Reload Default Factory Settings
- A.2.10 Wakeup Magnet
- A.2.11 Batteries
- A.2.12 Unit Dimensions
- A.2.13 Enclosure Material
- A.2.14 Environmental Ratings
- A.3 Kaye ValProbe RT Logger
- A.3.1 Installation Category
- A.3.2 Overvoltage Category and Electrical Safety Protection Class
- A.3.3 Magnetic Switch
- A.3.4 LED Indications
- A.3.5 Calibration
- A.3.6 Real Time Clock Accuracy
- A.3.7 Sensing Elements
- A.3.8 Housing Material
- A.3.9 Logger Base Dimensions
- A.3.10 Ingress Protection
- A.3.11 Insulating Canister X2545
- A.3.12 Battery
- A.3.13 Shelf Life
- A.3.14 Battery Life
- A.4.1 Measurement Accuracy
- A.4.2 Body Environmental Ratings:
- A.4.3 Number of Input Channels
- A.4.4 Sensor Types and Measurement Range
- A.4.5 Sensor Lengths
- A.5.1 Measurement Accuracy
- A.5.2 Body Environmental ratings:
- A.5.3 Number of Input Channels
- A.5.4 Sensor Types and Measurement Range
- A.5.5 Sensor Lengths
- A.6.1 Measurement Accuracy
- A.1 Kaye ValProbe RT System
- Appendix B. Regulatory Compliance and Agency Approvals
- B.1 Electrical Safety Approvals:
- B.2 Radio and EMC Certification Approvals:
- B.3 Ingress Protection for Kaye ValProbe RT Logger
- B.4 United States FDA 21 CFR Part 11 Complaint Software Application
- B.5 Transportation Regulations
- B.6 FCC Compliance Statement
- B.7 ISED Canada Compliance Statement
- B.8 EU Compliance Statement:
- B.9
- Product Certification Labels
- Appendix C. Environnemental Compliance
- Appendix D. Warranty and Returns:
- Appendix E. Service Information
Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual 90
Table 6: Recommended Temperature Baths
Model
Bath
Uncertainty
IRTD
Uncertainty
Maximum Total
Uncertainty
CTR-25 (-25°C to 140°C) – Liquid
0.020°C
0.025°C
0.045°C
CTR-40 (-40°C to 150°C) – Liquid
0.010°C
0.025°C
0.035°C
CTR-80 (-80°C to 30°C) – Liquid
0.012°C
0.025°C
0.037°C
LTR-150 (-30°C to 150°C) –
Liquid/Dry
0.10°C 0.025°C 0.125°C
Note: Dry Blocks cannot be utilized for Verification to manufacturers specifications (0.1°C) as
they do not have the necessary Uncertainty.
If your verification strategy is to do a periodic pre-Verification and post Verification of the
process qualification, then the Verification criteria is usually < 0.5°C. In this case the above-
mentioned baths or Kaye Dry Blocks can be utilized as they provide adequate uncertainty as
compared to the criteria.
Table 7: Recommended Temperature Dry Blocks
Model
Bath
Uncertainty
IRTD
Uncertainty
Maximum Total
Uncertainty
LTR-150 (-30
°C to 150°C) – Dry Block
0.10°C
0.025°C
0.125°C
LTR-40 / 140 (-40
°C to 150°C) – Dry
Block
0.10°C 0.025°C 0.125°C
LTR-90 (-90
°C to 140°C) – Dry Block
0.15°C
0.025°C
0..175°C
HTR 420 (30
°C
to 420
°C) – Dry Block
0.10
°C
0.025°C
0.125°C
HTR 400 (-25
°C
to 400
°C) – Dry Block
0.10
°C
0.025°C
0.125°C
LTR -25/140 (-25°C to 140°C) – Dry
Block
0.15°C 0.025°C 0.04°C
Note: The sensor tips of bendable or flexible loggers should be fully inserted to the bottom of
the well. Rigid loggers with shorter tips have to utilize a liquid bath.
Pressure Sensors
In order to verify the pressure sensors, meet the calibration accuracy specification, the
pressure calibration equipment must have an accuracy of 0.15% full scale (of 5 bar) or better.