Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual 99
The Verification Results screen will display the Pass / Fail analysis results of the Verification
for each sensor. You can select the setpoints to view the results for each setpoint.
Additional Verification report options:
Check detailed sensor readings and IRTD readings using the Sensor Readings
button.
Display the sensor and IRTD readings in graphical format
Generate a PDF Verification report for all selected verified sensors
Change various criteria, in the textboxes
When completed with the Verification Results select Exit.
10.6. Reasons for Verification Failure
Verification may fail for some sensors due to one of the following reasons:
Failed criteria of allowed Logger deviation
Failed Verification due to unstable IRTD and Loggers
Logger failed stability criteria and cannot be evaluated
Time of stability of IRTD and Logger may not overlap a sufficient time
Logger were removed from the bath/calibrator before the setpoint was reached
10.7. Verification Report
Once the process is completed, a Verification report can be generated
Figure 114: Verification Report