Chapter 9: Qualification Study
Kaye ValProbe RT User Manual 86
9.3. Viewing the Active Qualification Study
To start a study, press Start Qual or use the automatic start condition defined in the setup (if
Start Qual is not available) and enter User Credentials.
Figure 93: List View - Real-Time Readings
During the qualification study, real-time sensor readings and calculations for each group
defined in the setup are displayed. Inside the message window, real-time events that apply to
the whole study (start/stop qualification, start/stop exposure), as well as all group-specific
events are displayed.
The following guidelines apply when viewing sensor readings and calculations:
Calculations are computed, events are monitored and displayed on the Console screen.
The resolution of sensor readings and calculation results are dependent on the lowest resolution
in the group.
Open, under-range, over-range, or failed sensors are not allowed to participate in calculations or
events.
Each box lists the group name, sensor serial number, sensor label, and associated measurand
(temperature, pressure, etc.). Also listed are battery life, RF strength, and accumulated lethality if
defined in the setup. If there are more sensors or calculations that can fit on the screen, use the
scroll bar to display the remaining data.
A button to stop qualification is available at the bottom of the screen.
Note: The ValProbe RT continues the study after disconnecting.
To stop the study, press Stop Qual.
After stopping the study, Save Study becomes active. Please save the study to your console.
If a console is connected to a ValProbe RT Base Station with an unread study the Console
insists on reading the study before starting any new studies.