Model 310 Plus Series User’s Manual Pub# 5600-025-Rev-B
WEEE/RoHS Compliance Statement EU Directives WEEE and RoHS To Our Valued Customers: We are committed to being a good corporate citizen. As part of that commitment, we strive to maintain an environmentally conscious manufacturing operation. The European Union (EU) has enacted two Directives, the first on product recycling (Waste Electrical and Electronic Equipment, WEEE) and the second limiting the use of certain substances (Restriction on the use of Hazardous Substances, RoHS).
KDS Model 310 Plus Series Contents Specifications..........................................................................................................................2 General Safety Summary......................................................................................................3 Features ..................................................................................................................................4 Keypad Functions........................................................
KDS Model 310 Plus Series Specifications Model Model 310 Plus Syringe Size 1–250 microliter Electrical Rating US model 115V~, 0.1A CE model 230V~, 0.06A Fuse 5x20 mm, 250V~ slow blow, 0.1A US Model 100-120V~, 50/60Hz CE Model 200-240V~,50/60Hz Voltage Operating Range Drive Mechanism Microprocessor controlled stepper motor (1⁄2 microstep), leadscrew Force 2 lbs (minimum) Pusher Advance Per Half Step 1.
KDS Model 310 Plus Series General Safety Summary Please read the following safety precautions to ensure proper use of your syringe pump. To avoid potential hazards and product damage, use this product only as instructed in this manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. To Prevent Hazard or Injury: Use Proper Power Supply The pump is supplied with an approved power supply and line cord.
KDS Model 310 Plus Series Features The Model 310 Plus is an infusion/withdrawal syringe pump designed to hold glass syringes [Hamilton] from 1 to 250 microliters. Setup and pump operation is extremely simple. A menu, displayed on an alphanumeric LCD "prompts" the operator to make the necessary selections using the keypad for choice of features and numerical entries. The internal diameter of the syringe is used by the control program to calibrate the pump and deliver the volume and flow rate selected.
KDS Model 310 Plus Series Keypad Functions (continued) select This key has three functions: a. Return to main menu when pressed repeatedly b. Selects the pulsing or "highlighted" feature of the main menu display. c. Enter numerical values. a. Turns the motor on and off. b. Acts as a "pause" during a dispense. run/stop DISPLAY After the pump is programmed an arrow on the right side of the display indicates the direction of operation. This arrow pulses when the pump is running.
KDS Model 310 Plus Series Operating Instructions (Continued) SYRINGE DIAMETER ENTRY For automatic internal calibration the syringe diameter must be entered. Once entered the diameter is retained in non-volatile memory and need not be entered again unless the syringe used is changed. When a syringe diameter is changed the Volume and Rate settings are set to zero. There are two methods of diameter entry. a. Lookup Table Select TABLE from the main line menu.
KDS Model 310 Plus Series Operating Instructions (Continued) FLOWRATE ENTRY RATE can be selected from the main menu or will be displayed automatically after the Volume setting. The display reads: Rate: XX.XX µl/min 1) Use the arrow keys to display the flow rate required. 2) Enter the rate with the select key. 3) If the rate entered is out of range, the display will show the highest rate possible. To continue enter a lower rate, or select a larger syringe.
KDS Model 310 Plus Series Operating Instructions (Continued) CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING While the pump is running it is possible to look at the volume and flowrate settings without interrupting the operation of the pump. If required, the volume and flowrate settings can be changed while the pump continues to run. When the changes are entered the pump immediately changes to the new settings. 1) While operating, press select to return to the main menu.
KDS Model 310 Plus Series Loading The Syringe Raise the spring loaded syringe clamp and rotate away from the syringe barrel. Using the fast forward/fast reverse feature position the pusher block so that the syringe plunger and syringe barrel will fit into the syringe holder and pusher block. Place the syringe barrel in the V of the syringe holder, making sure that the flange of the syringe barrel is positioned between the two upright posts, and lower the syringe plunger knob into the pusher block.
KDS Model 310 Plus Series MAINTENANCE Maintenance is required only for the moving mechanical parts which should be kept clean and lubricated. Occasionally, a small amount of light machine oil should be applied to the guide rods and a small amount of grease or oil to the leadscrew. Solvents of any type should never be used to clean the pump. A mild detergent solution may be used to clean the keypad. Table 1 Flow Rate / Syringe Diameter Table: 310 Plus Syringe Min. µ/min Max. µ/min 0.5 µl 0.001 0.
KDS Model 310 Plus Series Limited Warranty KD Scientific Inc. warrants to the first consumer purchaser, for a period of one year from the date of purchase, that this unit, when shipped in its original container, will be free from defective workmanship and materials and agree that it will, at its option, either repair or replace the defective unit. This warranty does not extend to misuse, neglect or abuse, normal wear and tear, accident, modification or unauthorized repair.
WEEE/RoHS Compliance Statement EU Directives WEEE and RoHS To Our Valued Customers: We are committed to being a good corporate citizen. As part of that commitment, we strive to maintain an environmentally conscious manufacturing operation. The European Union (EU) has enacted two Directives, the first on product recycling (Waste Electrical and Electronic Equipment, WEEE) and the second limiting the use of certain substances (Restriction on the use of Hazardous Substances, RoHS).