Configuring the KDS Legato 200 Series
KDS Legato 200 Series User’s Manual
KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
S
etting the Pump Address
Up to 100 pumps can be linked together. Each pump
can be given a unique address.
I
n the Set-up Screen press
P
ump Address
.
Valid entries are 00 to 99.
Enter the pump address.
Accept the changes made by pressing
and go back to the Set-up screen.
C
ancel
t
he changes made by pressing
and go back to the Set-up screen.
Press Here
Press Here
Press here to
select Baud Rate
Setting the RS 232 Baud Rate
The Legato 200 Series has an RS 232 port for serial
communications.
To adjust the baud rate from
9600 to115,200: in the Set-up screen
press
Baud Rate
Select the Baud Rate.
Accept the changes made by pressing
and go back to the Set-up screen.
Cancel the changes made by pressing
and go back to the Set-up screen.