Owner manual
28
Configuring the KDS Legato 200 Series
KDS Legato 200 Series User’s Manual
KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
R
esume Run on Power Loss
If Resume Run on Power Loss is selected the pump will
power up running when power resumes.
In the Set-up screen,
select
Resume Run After Power Loss
Accept the changes made and go back
to the previous screen before Set-up mode
was selected.
Cancel and go back to the previous
screen before Set-up mode was selected.
✐
NOTE: This feature is only valid in Infuse
Only or Withdraw Only modes. This feature
will only execute if the pump is running at
t
he time when power is lost, and was
running with no target set.
P
ress Here
Press Here
Setting and Viewing Pump Information
The information on the pumps parameters are displayed
in the Pump Information Screen.
In the Set-up screen, select
Pump Information
The current information on the pump will be displayed.
• The Model type
• The Software Version
• The # of program steps available
• The Calibration Due Date
• The Maintenance Due
• The Current Pump Address
Press
Calibration Reset to reset the
calibration date (one year from current date).
Confirm the Calibration Reset by pressing the
Calibration Reset again.
Press Maintenance Reset to reset the # of hours
for maintenance. This will reset it to 100 hours.
Confirm the Maintenance Reset by pressing the
Maintenance Reset again.
Accept the changes made by pressing
and go back to the Set-up screen.
Cancel the changes made by pressing
and go back to the Set-up screen.
Press here to
reset the
maintenance
counter
Press here to
reset the
calibration date
phone 508.429.6809 fax 508.893.0160
e-mail info@kdscientific.com web www.kdscientific.com
✐
NOTE: Elapsed Timers and Total Infused
Volume will be reset when power resumes.
Counters and timers information prior to
loss of power are lost