Owner manual

60
Operating the KDS Legato 200 Series
KDS Legato 200 Series Users Manual
KD Scientific Legato 200 Series Syringe Pump Users Manual
Publication 5617-002-Rev F
ECHO
Sets or displays the RS232/USB echo state. Valid states
are on or off.
Command format:
echo [on|off]
Query response:
<lf>[##:]OFF<cr>
<lf>[##]<prompt>
or:
<lf>[##:]ON<cr>
<lf>[##]<prompt>
FREE
Displays the number of free program steps left on the
disk.
Command format:
free
Query response:
<lf>[##:]# steps used<cr>
<lf>[##:]# steps free<cr>
<lf>[##:]# total steps<cr>
<lf>[##]<prompt>
The number of steps displayed is a four digit number, right
justified, and padded with spaces.
FORCE
Sets or displays the infusion force level in percent. Valid
range is 1 to 100.
Command format:
force [1-100]
Query response:
<lf>[##:]#%<cr>
<lf>[##]<prompt>
FTSWITCH
Sets or displays the footswitch setting. The footswitch
may be used in a toggle on or off mode (momentary), a
press to run mode (falling), or a release to run mode
(rising).
Command format:
ftswitch [mom|rise|fall]
Query response:
<lf>[##:]Momentary<cr>
<lf>[##]<prompt>
or:
<lf>[##:]Active high<cr>
<lf>[##]<prompt>
or:
<lf>[##:]Active low<cr>
<lf>[##]<prompt>
POLL
Sets or displays the polling mode state. In polling mode,
prompts are inhibited if they aren’t generated by a typed
command. Valid states are on or off.
Command format:
poll [on|off]
Query response:
<lf>[##:]OFF<cr>
<lf>[##]<prompt>
or:
<lf>[##:]ON<cr>
<lf>[##]<prompt>
TILT
Calibrates the tilt sensor.
Command format:
tilt
Query response:
<lf>[##:]Calibrating the tilt sensor...<cr>
<lf>[##:]Tilt calibration value is #<cr>
<lf>[##]<prompt>