Operating the KDS Legato 200 Series
KDS Legato 200 Series Users Manual
phone 508.429.6809 fax 508.893.0160
e-mail info@kdscientific.com web www.kdscientific.com
KD Scientific Legato 200 Series Syringe Pump Users Manual
Publication 5617-002-Rev F
T
ime commands
CITIME
Clears the infused time. Configuration Run Screen
m
ode only.
Command format:
citime
CTIME
Clears both the infused and withdrawn times.
C
onfiguration Run Screen mode only.
Command format:
ctime
CTTIME
Clears the target time. Configuration Run Screen
mode only.
Command format:
cttime
CWTIME
Clears the withdrawn time. Configuration Run Screen
mode only.
Command format:
cwtime
ITIME
Displays the infused time. Configuration Run Screen
mode only.
Command format:
itime
Query response:
<lf>[##:]# seconds<cr>
<lf>[##]<prompt>
or:
<lf>[##:]##:##:##<cr>
<lf>[##]<prompt>
TTIME
Sets or displays the target time. Configuration Run Screen
mode only.
Command format:
ttime [{target time}]
Query response:
<lf>[##:]Target time not set<cr>
<lf>[##]<prompt>
or:
<lf>[##:]# seconds<cr>
<lf>[##]<prompt>
or:
<lf>[##:]##:##:##<cr>
<lf>[##]<prompt>
WTIME
Displays the withdrawn time. Configuration Run Screen
mode only.
Command format:
wtime
Query response:
<lf>[##:]# seconds<cr>
<lf>[##]<prompt>
or:
<lf>[##:]##:##:##<cr>