User manual

Delay
A programmable delay is available after event detection. It can be set manually or an auto
delay can be used. With auto delay, the Model 2000 selects a delay based on the function
and range. The AUTO settings are listed in Table 3-2.
The delay function is accessed by pressing the SHIFT-DELAY keys. The present delay
setting (AUTO or MANual) is displayed. Use the
and keys to select the type of delay.
If MANual is chosen, also enter the duration of the delay. The maximum is shown following:
99H:99M:99.999S
Press ENTER to accept the delay or EXIT for no change.
Changing the delay to MANual on one function changes the delays on all functions to
MANual.
Table 3-2
Auto delay settings
Function Range and delay
DCV
ACV
FREQ
100mV
1ms
100mV
400ms
100mV
1ms
1V
1ms
1V
400ms
1V
1ms
10V
1ms
10V
400ms
10V
1ms
100V
5ms
100V
400ms
100V
1ms
1000V
5ms
750V
400ms
750V
1ms
DCI
ACI
10mA
2ms
100mA
2ms
1A
2ms
1A
400ms
3A
2ms
3A
400ms
Ω2W, Ω4W
Continuity
Diode testing
100Ω
3ms
1kΩ
3ms
1kΩ
3ms
1mA
1ms
10kΩ
13ms
100µA
1ms
100kΩ
25ms
10µA
1ms
1MΩ
100ms
10MΩ
150ms
100MΩ
250ms
2000-900 (J - Aug 2010) BOOK.fm Page 9 Wednesday, October 12, 2011 12:29 PM