Buffer operations
The Model 2000 has a buffer to store from two to 1024 readings and units. It also stores
the channel number for scanned readings and overflow readings. In addition, recalled data
includes statistical information, such as minimum, maximum, average, and standard devia
-
tion.
The buffer fills with the requested number of readings and stops. Readings are placed in
the buffer after any math operations are performed. Buffered data is overwritten each time
the storage operation is selected. The data is volatile; it is not saved through a power cycle.
The following paragraphs discuss storing and recalling buffered data.
Storing readings
Use the following procedure to store readings:
1. Set up the instrument for the desired configuration.
2. Press the STORE key.
Using the , , , and keys to select the number of readings desired.
4. Press ENTER. The asterisk (*) annunciator turns on to indicate a data storage oper-
ation. It will turn off when the storage is finished.
2000-900 (J - Aug 2010) BOOK.fm Page 17 Wednesday, October 12, 2011 12:29 PM