User manual
IEEE-488 Reference
3-121
Bit B1, Sequence 1 (Seq1) —
Set bit indicates that the instrument is in an arm layer (PTR), or
that the instrument has exited from the arm layers (NTR).
Bits B2 through B15 —
Not used.
Sequence Event Register:
Bit B0 —
Not used.
Bit B1, Layer 1 (Lay1) —
Set bit indicates that instrument operation is in arm layer 1 (PTR),
or that operation has exited from arm layer 1 NTR).
Bit B2, Layer 2 (Lay2) —
Set bit indicates that instrument operation is in arm layer 2 (PTR),
or that operation has exited from arm layer 2 NTR).
Bits B3 through B15 —
Not used.
PRINT #1, "output 16; :stat:meas?" ' Query Measure Event Register
PRINT #1, "enter 16" ' Get response from 2002
Figure 3-25
Arm Event Register
B15 - B2
B1 B0
(2 )
1
2
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Event Bit Set
0 = Event Bit Cleared
Seq1
Event : Seq1 = Sequence 1
Program fragment
Figure 3-26
Sequence Event Register
Lay2
B15 - B3 B2 B1 B0
(2 )
2
4
(2 )
1
2
0/1 0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Event Bit Set
0 = Event Bit Cleared
Event : Lay1 = Layer 1
Lay2 = Layer 2
Lay1