Model 2002
Multimeter Specifications
Resistance Uncertainty: = ±[(ppm of reading) x (measured value) + (ppm of range) x (range used)] / 1,000,000.
% Accuracy: = (ppm accuracy) / 10,000.
1ppm of Range: = 20 counts for ranges up to 200M and 10 counts on 1G range at 7H digits.
SPEED AND ACCURACY 90 Days
Accuracy
11, 36
±(ppm of reading+ppm of range+ppm of range rms noise
12
)
RANGE
10PLC DFILT On,
10 Readings
10PLC
DFILT Off
1PLC DFILT On,
10 Readings
1PLC
DFILT Off
0.1PLC
37
DFILT Off
0.01PLC
37, 13
DFILT Off
20 15+11+0 15+11+0.5 15+13+0.5 15+13+1 15+16+25 110+200+35
200 15+8+0 15+8+0.5 17+8+0.5 17+8+1 17+10+15 110+200+35
2 k 7+0.8+0 7+0.8+0.05 8+0.8+0.07 8+0.8+0.2 8+1+2 130+230+5
20 k 7+0.8+0 7+0.8+0.1 8+0.8+0.1 9+0.8+0.2 40+1+2 130+230+5
200 k 29+0.8+0 29+0.8+0.1 31+0.8+0.1 34+0.8+0.2 250+1+2
2 M 55+0.5+0 53+0.5+0.1 58+0.5+0.1 68+0.5+0.2 750+0.7+2
20 M
32
175+0.6+0 175+0.6+0 175+0.6+0 200+0.6+0
200 M
32
500+3+0 510+3+0 510+3+0 550+3+0
1 G
32
2000+15+0 2100+15+0 2100+15+0 2500+15+0
PLC = Power Line Cycles. DFILT = Digital Filter.
2-WIRE ACCURACY ±(ppm of range)
Range
Additional
Uncertainty
(inside T
CAL
± 5°C)
Temperature
Coefficient
(outside T
CAL
±5°C)
20 300 ppm 70 ppm/°C
200 30 ppm 7 ppm/°C
2 k
3 ppm 0.7 ppm/°C
NORMAL ACCURACY
36
1PLC, Offset comp. off, DFILT off
Relative Accuracy
± (ppm of reading + ppm of range)
RANGE 24 Hours
35
90 Days
7
1 Year
7
2 Years
7
Temperature Coefficient
± (ppm of reading + ppm of range)/°C
Outside TCAL ± 5°C
20 5+12 15+16 17+17 20+19 2.5+2.5
200 7+8 17+11 19+12 22+13 2.5+1.8
2k 3.5+1.1 8+1.4 10+1.5 12+1.6 0.8+0.18
20k 4.5+1.1 9+1.4 11+1.5 13+1.6 0.8+0.18
200k 11+1.1 34+1.4 40+1.5 45+1.6 3.5+0.18
2M 27+0.9 68+1.1 80+1.1 90+1.1 7+0.1
20M
32
75+0.2 200+0.6 275+0.6 325+0.6 20+0.1
200M
32
200+1 550+3 600+3 650+3 80+0.5
1G
32
1250+5 2500+15 2550+15 2600+15 400+2.5
Settling Characteristics: Pre-programmed settling delay times are for <500pF external circuit capacitance. Reading settling times are affected
by source impedance and cable dielectric absorption characteristics.
Ohms Voltage Drop Measurement: Available as a multiple display.
Autoranging: Autoranges up at 105% of range, down at 10% of range.
2-WIRE RESISTANCE READING RATES
12, 17
Readings/Second to Memory Readings/Second to IEEE-488
18
Readings/Second with
Time Stamp to IEEE-488
18
PLC
Measurement
Aperture Bits
Default
Digits
Autozero Off Auto zero On Autozero Off Autozero On Autozero Off Autozero On
10 167 ms (200 ms) 29 6 ( 5 ) 2 (1.7) 6 (5) 2 (1.6) 6 (5) 2 (1.6)
2 33.4 ms (40 ms) 27 29 (25) 9 (7.6) 29 (24) 9 (7.4) 27 (22) 9 (7.4)
1 16.7 ms (20 ms) 26 56 (48) 47 (40) 55 (45) 46 (38) 50 (41) 42 (34)
0.2
37
3.34 ms (4 ms) 23 222 (197) 156 (139) 220 (196) 148 (132) 156 (139) 107 (95)
0.1
37
1.67 ms (2 ms) 22 330 (317) 176 (169) 305 (293) 166 (159) 157 (151) 110 (106)
0.02
37
334 µs (400 µs) 20 330 (330) 182 (182) 305 (305) 172 (172) 160 (160) 113 (113)
0.01
37
167 µs (167 µs) 19 384 (384) 186 (186) 352 (352) 172 (172) 179 (179) 123 (123)
0.01
37
167 µs (167 µs) 19 2000 (2000) 2000 (2000)
36
Specifications are for 1 power line cycle, normal autozero, digital filter off, autorange off, 4-wire mode, offset compensation off, except as noted.
37
Ohms measurements at rates lower than 1 power line cycle are subject to potential noise pickup. Care must be taken to provide adequate shielding.