User manual

HP3458A Emulation Mode
H-13
Description The NRDGS command programs the number of readings per trigger as well as the event that initiates
the trigger. Available events include continuous, external trigger, the GPIB talk command, as well as
timer events.
All other HP3458A trigger events are not supported by the Model 2002.
Example NRDGS 10, SYN Take 10 readings per talk command.
NRDGS 100, EXT Take 100 readings per external trigger event.
OCOMP
Purpose To enable or disable the offset-compensated ohms mode.
Format OCOMP <control>
Parameter <control>
<control> parameter Numeric equivalent Description
OFF 0 Offset-compensated ohms off
ON 1 Offset-compensated ohms on
Default ON (power-on default = OFF)
Query OCOMP? Returns the current offset-compensated ohms state.
Description The OCOMP command enables/disables the offset-compensated ohms function.
Example OCOMP ON Enable offset-compensated ohms.
OFORMAT
Purpose To program the data format for both normal and memory readings.
Format OFORMAT <format>
Parameter <format>
<format> parameter Numeric equivalent Description
ASCII 1 ASCII
SREAL 4 Single real
DREAL 5 Double real
Default ASCII
Query OFORMAT? Returns the present setting of the output format.