HP3458A Emulation Mode
H-15
FAST
Same as NORM with the following exceptions:
DCV 10 OFORMAT DINT
DISP OFF TRIG AUTO
AZERO OFF
DIG: Not supported.
Default NORM
Description The PRESET command selects one of two preset instrument configurations, one for normal opera-
tion, and one for fast operation. The DIG preset configuration supported by the HP3458A is not sup-
ported by the Model 2002.
Example PRESET FAST Select fast configuration.
RANGE
Purpose To select the measurement range.
Format RANGE <max_input> ,<%_resolution>
Parameter <max_input> Manual range or AUTO (autorange)
<%_resolution> Ignored
Default AUTO
Query RANGE? Returns the present measurement range for the current function.
Description The RANGE command allows you to select a manual measurement range or autorange. (See paragraph
3.18 for range information specific to each function.) Specifying AUTO for the <max_input>
parameter enables autoranging, and the <%_resolution> parameter is ignored.
Example RANGE AUTO Select autorange.
RESET
Purpose To reset the Model 2002 to the power-on state.
Format RESET
Parameter None
Description The RESET command resets the multimeter to the power-on state without cycling power. Reset
effectively executes: