HP3458A Emulation Mode
H-18
<type> parameter Numeric equivalent Description
ANA 1 RMS conversion
RNDM 2 Low-frequency voltage measurements
SYNC 3 Low-frequency voltage measurements
Default ANA
Query SETACV? Returns the currently specified conversion technique. A “3” is returned for RNDM
or SYNC.
Description The SETACV command selects the RMS conversion technique to be used for the AC or AC+DC volt-
age measurement modes. Note that both RNDM and SYNC choose the low-frequency voltage mea-
surement mode.
Example SETACV SYNC Select low-frequency AC mode.
SRQ
Purpose To enable SRQ.
Format SRQ
Parameter None
Description The SRQ command sets bit 2 in the Model 2002 status register. When bit 2 is enabled, the SRQ line
will be asserted when certain conditions occur. (See RQS command.)
SSTATE
Purpose To store the instrument's present configuration in memory.
Format SSTATE <state>
Parameter <state>
Memory option Valid <state> parameters
No extended memory SAV0
2002-MEM1 installed SAV0-SAV4
2002-MEM2 installed SAV0-SAV9
Description SSTATE stores the multimeter’s present state in the given state memory location. Note that the num-
ber of available states depends on installed memory options as indicated above. Instrument states
may recalled with the RSTATE command.
Example SSTATE SAV0 Save state 0.