User manual

HP3458A Emulation Mode
H-21
TEMP?
Purpose To request the internal temperature of the Model 2002.
Format TEMP?
Response Model 2002 internal temperature in °C
Description The TEMP? query returns the internal temperature of the Model 2002. The response is in standard
numeric format in degrees Celsius.
TERM
Purpose To request INPUTS switch status.
Format TERM <source>
Parameter <source> Not used
Query TERM? Returns status of INPUTS switch (1= Front; 2=Rear)
Description The TERM command is a no-op, and thus it is ignored. It is included for HP3458A compatibility. The
TERM? query returns the status of the INPUTS switch as indicated above.
TIMER
Purpose To define the timer sample interval.
Format TIMER <time>
Parameter <time> 0s to 6000s
Default 1s (power-on = 0s)
Query TIMER? Returns the currently specified time interval in seconds.
Description The TIMER command defines the time interval in seconds for the TIMER sample event in the NRDGS
command. The Model 2002 will accept time values from 0s to 6000s, but will modify values appro-
priately for compatibility.
Example TIMER 2 Define 2 second sample interval.