User Manual
Appendix B: Calibration  Series 2200 Programmable DC Power Supplies Reference Manual 
B-4  2200S-901-01 Rev. A/November 2011 
DC voltage accuracy without remote sense 
Instrument 
DUT voltage 
Test current 
Min 
Measured 
Max 
0% Test Voltage 
2200-20-5  
0 V 
2.5 A 
-0.0030 V 
__________ 
0.0030 V 
2200-30-5  
0 V  
2.5 A 
-0.0030 V 
__________ 
0.0030 V 
2200-32-3  
0 V 
1.5 A 
-0.0030 V 
__________ 
0.0030 V 
2200-60-2  
0 V 
1.25 A 
-0.0060 V 
__________ 
0.0060 V 
2200-70-1  
0 V 
0.63 A 
-0.0060 V 
__________ 
0.0060 V 
25% Test Voltage 
2200-20-5  
5 V 
2.5 A 
4.9955 V 
__________ 
5.0045 V 
2200-30-5  
7.5 V 
2.5 A 
7.49475 V 
__________ 
7.50525 V 
2200-32-3  
8 V 
1.5 A 
7.9946 V 
__________ 
8.0054 V 
2200-60-2  
15 V 
1.25 A 
14.9895 V 
__________ 
15.0105 V 
2200-70-1  
18 V 
0.63 A 
17.9886 V 
__________ 
18.0114 V 
50% Test Voltage 
2200-20-5  
10 V 
2.5 A 
9.9940 V 
__________ 
10.0060 V 
2200-30-5  
15 V 
2.5 A 
14.9925 V 
__________ 
15.0075 V 
2200-32-3  
16 V 
1.5 A 
15.9922 V 
__________ 
16.0078 V 
2200-60-2  
30 V 
1.25 A 
29.9850 V  
__________ 
30.0150 V 
2200-70-1  
36 V 
0.63 A 
35.9832 V 
__________ 
36.0168 V 
75% Test Voltage 
2200-20-5  
15 V 
2.5 A 
14.9925 V 
__________ 
15.00750 V 
2200-30-5  
22.5 V 
2.5 A 
22.49025 V 
__________ 
22.50975 V 
2200-32-3  
24 V 
1.5 A  
23.9898 V 
__________ 
24.0102 V 
2200-60-2  
45 V 
1.25 A 
44.9805 V 
__________ 
45.0195 V 
2200-70-1  
54 V 
0.63 A 
53.9778 V 
__________ 
54.0222 V 
100% Test Voltage 
2200-20-5  
20 V 
2.5 A 
19.9910 V 
__________ 
 20.0090 V 
2200-30-5  
30 V 
2.5 A 
29.9880 V 
__________ 
 30.0120 V 
2200-32-3  
32 V 
1.5 A 
31.9874 V 
__________ 
 32.0126 V 
2200-60-2  
60 V 
1.25 A 
59.9760 V 
__________ 
 60.0240 V 
2200-70-1  
72 V 
0.63 A 
71.9724 V 
__________ 
 72.0276 V 