User Manual
Series 2200 Programmable DC Power Supplies Reference Manual  Appendix B: Calibration 
2200S-901-01 Rev. A/November 2011  B-13 
Item 
Connectors 
Minimum requirements 
Example 
2200-20-5 
5  
HL10006E5R000JJ 
2200-30-5 
10  
HL10006E10R00JJ 
2200-32-3 
12  
HL10006E12R00JJ 
2200-60-2 
40  
HL10006E40R00JJ 
2200-72-1 
65  
HL10006E65R00JJ 
11. 
Load Resistor for Current 
Noise testing 
1 , 100 W, 5% 
To perform the tests as illustrated, both leads of the 
load resistor require a Banana plug/jack (or equivalent) 
to connect to the power supply. 
Vishay/Dale 
HL10006Z1R000J
J 
12. 
Oscilloscope 
20 MHz bandwidth limited oscilloscope at 1 
mV/division 
Tektronix 
DPO3012 
13. 
Oscilloscope probe 
Low capacitance 1 M/10 M 1X/10X 6 MHz/200 
MHz probe 
Tektronix P2220, 
1X probe 
14. 
Oscilloscope probe 
Low capacitance 10 M 500 MHz probe 
Tektronix P6139A, 
10X probe 
15. 
Coaxial cable 
50  BNC, male-to-male 
Tektronix part 
number: 012-
0482-00 
16. 
BNC adapter 
BNC female to banana breakout 
Pomona 
Electronics 3073 
Self test 
This procedure uses internal routines to verify that the DUT functions and passes its internal self 
tests. No test equipment or hookups are required. 
1.  Disconnect all cables from the DUT outputs. 
2.  Power on the DUT. The front-panel display will light up briefly while the DUT performs its power-
on self test. All the display annunciators will light up at once. 
3.  Review the display with all the annunciators. Visually check if there are any strokes lost on any 
annunciator. 
Figure 17: Display 