User Manual
Section 2: General operation  Series 2200 Programmable DC Power Supplies Reference Manual 
2-30  2200S-901-01 Rev. A/November 2011 
Timer example 
This example sets up the power supply to output 6 V, 1 A for 7 seconds before the output 
automatically turns off. 
Keystrokes 
Perform the following procedure using the keys on the instrument's front panel to save the test setup. 
To configure the duration of the timer and enable the timer feature: 
1.  Press Shift and Menu (1) to access the configuration menus. 
2.  Use the arrow keys, navigation wheel  , or the keypad to select Protect. Press Enter. 
3.  Use the arrow keys, navigation wheel  , or the keypad to select Out Time Set. Press Enter. 
4.  Use the arrow keys, navigation wheel  , or the keypad to select On. Press Enter. 
5.  Use the arrow keys, navigation wheel  , or the keypad to input Time = 7.00 S. Press 
Enter. 
6.  Press Esc a few times to return to the initial screen. 
7.  Press V-Set.  
8.  Use the arrow keys, navigation wheel  , or the keypad to input 6 V. Press Enter. 
9.  Press I-Set. 
10. Use the arrow keys, navigation wheel  , or the keypad to input 1 A. Press Enter. 
11. Press Output On/Off to turn on the output. 
Wait for 7 seconds. The output will turn off automatically 
Programming commands 
*RST 
SOURCE:FUNC:MODE FIX 
SOURCE:VOLT 6V 
SOURCE:CURR 1A 
To configure the duration of the timer: 
OUTPUT:TIMER:DELAY 7S 
To enable the timer feature: 
OUTPUT:TIMER:STATE ON 
OUTPUT:STATE ON 
To disable the timer feature: 
OUTPUT:TIMER:STATE OFF 