Series 2200 Programmable DC Power Supplies Reference Manual  Section 4: Command reference 
2200S-901-01 Rev. A/November 2011  4-5 
FETCh:VOLTage[:DC]? (query only) 
This command returns the last measured output voltage stored in 
the communications buffer of the power supply. A new 
measurement is not initiated by this command. 
Group  
Measurement 
Syntax  
FETCh:VOLTage[:DC]? 
Related 
Commands  
MEASure:VOLTage[:DC]? (query only) (on page 4-7) 
Returns 
<NR2> is the measured output voltage in volts. 
Examples 
FETC:VOLT? might return 5.0011, which would be the measured 
voltage across the power supply outputs in volts. 
FETCh[:SCALar]:POWer? (query only) 
This command returns the last measured output current stored in 
the communications buffer of the power supply. A new 
measurement is not initiated by this command. The power 
calculation in the instrument is performed approximately every 100 
ms. Insure that the voltage and current are stable longer than this 
for good results. 
Group 
Measurement 
Syntax  
FETCh[:SCALar]:POWer? 
Returns 
<NR2> is the measured output power in watts. 
Examples 
FETCh:POW? might return 6.01667, which would be the power 
measured at the output of t he power supply in watts. 