User Manual
Series 2200 Programmable DC Power Supplies Reference Manual  Section 4: Command reference 
2200S-901-01 Rev. A/November 2011  4-7 
MEASure:VOLTage[:DC]? (query only) 
This command initiates and executes a new voltage measurement, 
and returns the measured output voltage of the power supply. 
Group 
Measurement 
Syntax  
MEASure:VOLTage[:DC]? 
Related 
Commands 
FETCh:VOLTage[:DC]? (query only) (on page 4-5) 
Returns 
<NR2> is the measured output voltage in volts. 
Examples 
MEAS:VOLT? might return 5.0011 which would be the voltage 
measured across the power supply output in volts. 
*OPC 
This command configures the instrument to generate an operation 
complete message by setting bit 0 of the Standard Event Status 
Register (SESR) when all pending commands that generate an 
OPC message are complete. 
The query command places the ASCII character "1" into the output 
queue when all such OPC commands are complete. 
Group 
Synchronization 
Syntax  
*OPC 
*OPC? 
Examples 
*OPC? might return 1 to indicate that all pending OPC operations 
are finished. 