[SOURce:]VOLTage:PROTection[:LEVel] 
This command sets the over voltage protection (OVP) threshold 
level. 
Group 
Source 
Syntax  
[SOURce:]VOLTage:PROTection[:LEVel] 
{<voltage>|MIN|MAX|DEF} 
[SOURce:]VOLTage:PROTection[:LEVel]? 
Related 
Commands 
[SOURce:]VOLTage:PROTection[:LEVel] (on page 4-26) 
[SOURce:]VOLTage:RANGe (on page 4-28) 
Arguments  
<voltage>::=<NRf><units> 
where: 
<NRf> is a flexible decimal that specifies the OVP threshold, 
ranging from 0 to the maximum OVP level 
<units> ::={V|mV} 
MIN: sets the minimum OVP level (1 V). 
MAX: sets the maximum OVP level, which is approximately 10% 
higher than the maximum output voltage rating of the power supply. 
DEF: sets the OVP level to the default level, which is equal to MAX. 
Returns 
<NR2> is the OVP threshold in volts. 
MIN: set to minimum OVP level. 
MAX: set to maximum OVP level. 
Examples 
VOLT:PROT 30V 
VOLT:PROT? might return 30, which would be the OVP threshold in 
volts. 