User manual

OPERATION
45
Clear OVP/OCP
protection
The OVP or OCP protection can be
cleared after it has been tripped by
holding the OVP/OCP button for 2
seconds.
(Only applicable when the power
switch trip setting is disabled
[F-95 = 1])
(hold)
Set to CV Mode
When setting the power supply to constant voltage mode, a current
limit must also be set to determine the crossover point. When the
current exceeds the crossover point, the mode switches to CC mode.
For details about CV operation, see page 18. CC and CV mode have
two selectable slew rates: High Speed Priority and Slew Rate
Priority. High Speed Priority will use the fastest slew rate for the
instrument while Slew Rate Priority will use a user-configured slew
rate.
Background
Before setting the power supply to CV mode,
ensure:
The output is off.
The load is connected.
Steps
1. Press the Function key. The
Function key will light up.
2. The display should show F-
01 on the top and the
configuration setting for
F-01 on the bottom.
3. Rotate the Voltage knob to change
the F setting to F-03 (V-I Mode
Slew Rate Select).
Voltage