User manual

2260B Series User Manual
58
OVP/ OCP
The master unit can shut down slave units
when OVP/OCP is tripped on the master
unit (if the slave connector is wired for shut
down on alarm).
OVP/OCP can be independently tripped on
each slave unit, however the shutdown of the
power or output of the unit is disabled. Only
the alarm will be enabled.
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) are only supported on
the master unit.
The IMON current represents the total
current of the all the parallelized units.
Remote Sense
Please see the remote sense chapter for
details, page 53.
External Voltage and Resistance Control
Voltage/Resistance controlled remote
control can only be used with the master
unit.